Paper
27 December 1977 Methods And Results Of Dynamic Range Testing Of Diagnostic Ultrasound Instrumentation
S. W. Smith, H. Lopez, H. F. Stewart
Author Affiliations +
Proceedings Volume 0127, Application of Optical Instrumentation in Medicine VI; (1977) https://doi.org/10.1117/12.955944
Event: Application of Optical Instrumentation in Medicine VI, 1977, Boston, United States
Abstract
The technological advance of gray scale imaging has been a major factor in the rapid growth of the clinical application of diagnostic ultrasound. Gray scale imaging implies that the displayed brightness of a target in the ultrasound image is a function of the amplitude of its echo. Several techniques have been investigated for the measurement of the dynamic range of gray scale ultrasound scanners. The measurement methods include (1) an electronic pulse burst generator which emits a sequence of pulses decaying at a selectable rate of 0, 1.25 or 2.5 dB/cm, (2) a polypropanol wedge designed to attenuate acoustic echoes over a range of 0 to 60 dB, (3) a series of stainless steel reflectors varying in angle from 0 to 10° so that returning echoes vary over a 40 dB range. All three instruments are easy to use. However, the wedge and variable angle devices must be calibrated for use with individual transducers. The electronic test device excludes the transducer when testing system dynamic range. Therefore, the pulse burst generator seems to be the most suitable for clinical or survey applications at this time.
© (1977) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. W. Smith, H. Lopez, and H. F. Stewart "Methods And Results Of Dynamic Range Testing Of Diagnostic Ultrasound Instrumentation", Proc. SPIE 0127, Application of Optical Instrumentation in Medicine VI, (27 December 1977); https://doi.org/10.1117/12.955944
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Transducers

Ultrasonography

Calibration

Scanners

Diagnostics

Medicine

Reflectors

RELATED CONTENT

Angioplasty under ultrasound?
Proceedings of SPIE (July 01 1990)
Gray Scale Imaging With Complex TGC And Transducer Arrays
Proceedings of SPIE (March 25 1976)
Overview Of Ultrasonic Imaging
Proceedings of SPIE (November 01 1982)
Practical Aspects Of Grayscale Ultrasound
Proceedings of SPIE (March 25 1976)

Back to Top