Paper
17 May 1989 Removing The Instrument Function From Fluorescence Spectra
Andrew F. Childs
Author Affiliations +
Proceedings Volume 1054, Fluorescence Detection III; (1989) https://doi.org/10.1117/12.951538
Event: OE/LASE '89, 1989, Los Angeles, CA, United States
Abstract
The spectrum acquired at the sample phototnultiplier tube of a fluorescence spectrophotometer is a product of the sample spectrum and the instrument function. The determination of the instrument function and its removal from the acquired spectrum is often critical to the accurate determination of the physical properties of the sample. Methods are discussed for the determination and removal of the instrument function from excitation and emission spectra. Methods considered include quantum counters and ratio circuits for excitation correction, and emission correction against calibrated excitation systems, calibrated tungsten lamps, and NBS standard quinine sulfate.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrew F. Childs "Removing The Instrument Function From Fluorescence Spectra", Proc. SPIE 1054, Fluorescence Detection III, (17 May 1989); https://doi.org/10.1117/12.951538
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Luminescence

Lamps

Tungsten

Calibration

Fluorescence spectroscopy

Spectrophotometry

Photomultipliers

RELATED CONTENT


Back to Top