Presentation
15 February 2021 Photon counting CT in a C-arm interventional system: hardware development and artifact corrections
Author Affiliations +
Abstract
This work reports the development of a C-arm photon counting detectors (PCD)-CT system for evaluating the potential clinical utility of PCD-CT for intraoperative imaging. A dual-threshold CdTe-based strip PCD was mounted on a Siemens Artis Zee C-arm interventional system. A new geometric calibration method was developed to correct for the geometric distortion of the C-arm system during rotation. Experimental results show that, under clinically relevant conditions, the C-arm PCD-CT system can reliably and reproducibly generate high quality MDCT-like images without any noticeable geometric distortion or banding artifact.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xu Ji, Mang Feng, Ran Zhang, Guang-Hong Chen, and Ke Li "Photon counting CT in a C-arm interventional system: hardware development and artifact corrections", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 115950A (15 February 2021); https://doi.org/10.1117/12.2581087