1 May 1994 Development of a technique for electronic assessment of the image quality of medical fluoroscopic systems
Henry Weatherburn, G. Manson
Author Affiliations +
Abstract
While methods of testing fluroscopic systems are well established, using test objects such as the Leeds test objects, observer influence may affect the results and the process is relatively time consuming. An alternative approach is examined, whereby use of a line selector and oscilloscope are used with a test object to examine resolution, contrast and grey scale steps. Three radiotherapy simulator fluoroscopic imaging systems are tested and, in each case, observed images are found to be limited by the signal response, as present in the video signal, and not by monitor performance. Edge enhancement effects and signal droop are also found to be effects which merit further investigation. It appears that, at low cost, the use of an oscilloscope and a TV line selector may permit the development of quantitative methods for assessing the performance of imaging systems.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Henry Weatherburn and G. Manson "Development of a technique for electronic assessment of the image quality of medical fluoroscopic systems", Proc. SPIE 2163, Medical Imaging 1994: Physics of Medical Imaging, (1 May 1994); https://doi.org/10.1117/12.174252