13 March 2009 Nonuniform angular dose distribution in digital breast tomosynthesis for increased conspicuity of small high contrast objects
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 72580Q (2009) https://doi.org/10.1117/12.813744
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Digital breast tomosynthesis (DBT) has been shown to decrease breast structural noise thus improving the detection of masses. However decreased detectability of microcalcifications was observed, and several studies have been performed to investigate the benefit of taking an additional central projection view after a DBT scan. Our study investigates the effect of variable angular dose distribution within a single DBT scan. Using a prototype DBT system with uniform angular dose distribution, several DBT scans (25 projection views over 40 degree angular range) were performed using different glandular doses. A subset of projection images was selected from each scan to form composite DBT scans (25 views each) with different angular dose distribution schemes (ADS). Two examples of ADS were: 1) seven central views with four times the dose of periphery; and 2) five central projections with six times the dose of periphery. The total dose for each ADS was the same as a reference scan with uniform dose distribution (1.5 mGy). They also had the same number of views and angular range, and were reconstructed using identical reconstruction filter settings. The detectability of calcifications, the 3D MTF and NPS of the system, and the ideal observer object detectability index for all cases were compared. The results showed that higher dose for the central views improve the detectability of calcifications. However magnitude of improvement depends on the reconstruction method and the size of the object.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yue-Houng Hu and Wei Zhao "Nonuniform angular dose distribution in digital breast tomosynthesis for increased conspicuity of small high contrast objects", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72580Q (13 March 2009); https://doi.org/10.1117/12.813744
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Digital breast tomosynthesis

Advanced distributed simulations

Signal to noise ratio

Image filtering

3D modeling

Modulation transfer functions

Electronic filtering

Back to Top