|
I.IntroductionThe Laser Interferometer Space Antenna (LISA) [1] aims to detect gravitational waves in the frequency range from 0.1 mHz to 1 Hz. The mission will consist of three satellites that orbit the sun in a heliocentric orbit, following the earth. Each satellite will carry free-flying test masses that define the endpoints of the interferometer arms. Laser light transmitted between the satellites will be used to read out the armlength changes with picometre/ Currently, an elegant breadboard of the LISA OB is in development by a consortium consisting of EADS Astrium GmbH – Satellites as prime contractor to the European Space Agency (ESA), TNO Science & Industry, University of Glasgow and Albert Einstein Institute [3]. Although this project was initiated while LISA was the L1 candidate mission, the basic metrology functions implemented and the results are also applicable to NGO/eLISA [4], [5]. Figure 1 shows the layout of the OB. The TX laser injects about 1.6W of light at 1064 nm wavelength on the left hand side. The main fraction of this light will be sent to the telescope and from there to the distant satellite. A fraction of the light is used for local interferometer readouts and another fraction is transmitted to the second OB on the same satellite. In the same way, light from the other OB on the satellite is injected to the OB on the top in Fig. 1 via the so-called backlink fiber. On the OB this so-called LO light is used with the TX light in the reference, PAAM, and test mass interferometer (in the upper left, lower left, and upper right in the figure, respectively). The science interferometer (in the lower right) interferes light from the distant satellite (RX light) with local TX light. Heterodyne interferometry will be used with heterodyne frequencies in the range from 2 to 20 MHz. Characterisation of the OB requires to externally close the test mass interferometer and the science interferometer by simulators and read-out of the respective photo detectors with a phase measurement system. In the following sections we give an overview of the experimental setup, the phase measurement system and the test mass and telescope simulator. II.Experimental SetupFigure 2 shows a simplified schematic of the experimental setup that will be used to test the performance of the science interferometer. Optical bench as well as telescope and test mass simulator will be placed inside the vacuum chamber (the simulator will be placed on top of the optical bench through Zerodur legs for pathlength stability). A combination of scroll pump and turbo pump is connected via diaphragm bellows to the chamber and keeps the pressure at a level below 10−5 mbar. The chamber rests on a self-levelling vibration isolation system that is also used for optical tables. It decouples the chamber from vibrations caused by e.g. the vacuum pumps. Laser LTX is a commercial nonplanar ring oscillator laser (Prometheus 20 by Innolight), that generates about 700mW of output power at 1064 nm and uses a nonlinear crystal in single-pass transmission to generate laser light of a few mW at the second harmonic frequency. This green light is used for frequency stabilisation of the laser to molecular iodine using modulation-transfer spectroscopy [6]. The laser output is amplified by an Ytterbium-doped fiber amplifier (PSFA-1064-01-5W-1-3 by Nufern). About 1.6W of optical power will be injected to the optical bench. Laser LR is an identical frequency stabilised laser system (without amplifier) that is used for frequency noise characterisation. The differential frequency noise between both stabilised lasers is about Fig. 2.Schematic layout of the test setup with optical bench, test mass and telescope simulator, lasers, temperature measurement system, and phase measurement system. ![]() The output of laser LBL is transmitted to the optical bench and is required for operation of reference, test mass and PAAM interferometers. When used, the frequency of laser LBL will be offset-phase-locked to laser LTX to a difference frequency in the 2 to 20MHz range. For operation of the science interferometer lasers LTX and LRX are required. Laser LRX is then transmitted to the telescope simulator. The signal from the transponder interferometer is used to offset-phase-lock laser LRX to laser LTX. In the beam path of laser LRX up to three mirror pairs can be inserted to attenuate the beam power. Mirror pairs with 45° angle of incidence were chosen because the beam direction and location is not altered upon insertion of the attenuators. A fraction of the power transmitted to optical bench and telescope simulator will be split off and measured with photodetectors. Their signals will be used to stabilise the laser output powers by feeding back to the fiber amplifier pump current and the laser pump current, respectively. The photodetector signals of the interferometers in use will be input to the phase measurement system, also called phasemeter, that is described in Sec. III. As auxiliary data, temperature sensors will be placed at different locations and the pressure in the vacuum chamber and the difference frequency between both stabilised lasers can be recorded. Figure 3 shows a schematic side view of the vacuum chamber. It will contain the optical bench with the telescope simulator on top. They will be surrounded by a thermal shield. The shield consists of walls of sheet Aluminum on an Aluminum baseplate. The optical bench and the thermal shield baseplate will rest on ceramic spacers made of Macor. Shield and spacer act as thermal low pass for heat conduction. The shield acts as a thermal capacitor, the spacers as thermal resistors. The metal shield is completely covered by multilayer insulation (ten pairs of reflecting foil and spacer) to increase the thermal resistance for thermal radiation. The thermal shield is surrounded by an intermediate mass made of Aluminum resting on Macor spacers. All cables connected to optical bench or simulator will be brought into contact with the intermediate mass before they are connected to feedthroughs in the vacuum chamber walls. The intermediate mass reduces temperature fluctuations of the cables originating from the laboratory environment. Figure 4 shows the temperature noise in the vacuum chamber at different locations. During the temperature noise measurement, the LISA optical bench was replaced by an optical bench that was used for interferometric actuator characterisation. The trace labelled ‘goal’ shows the temperature noise nT(f) we aim for during LISA optical bench testing, which is given by where uT(f) is given by Two different types of temperature sensors were used: Negative temperature coefficient thermistors (NTCs, dashed lines) and platinum sensors (PT, solid lines) both with resistances of 10 kΩ at 25 °C. The traces labelled ‘read-out limit’ were generated from the differences of two sensors each. They show the uncorrelated noise between these two channels and are a measure for the sensitivity of the temperature measurement system. With PT and NTC sensors, temperature noise floors of III.Phase Measurement SystemInitially we planned to use the phase measurement system (PMS) of LISA Pathfinder [7] for input signal frequencies in the kHz range. This would have required to mix the heterodyne signals generated on the optical bench from MHz frequencies to kHz frequencies. However, such a downmixing increases the phase readout noise by a factor of Fig. 5.Mixing of noise to interferometer heterodyne frequency. White noise present at the mirror frequency is folded to the downmixed signal, increasing its noise level by a factor of ![]() Figure 7 shows the schematic of one channel of the LISA PMS [8]. Externel to the PMS, two laser beams are interfered and detected by a photodiode. The current of the photodiode is converted to a voltage by the transimpendance amplifier (TIA) and filtered by an anti-aliasing filter (AAF) before being digitised by an analog-to-digital converter (ADC). The digital signal is mixed with sine and cosine local oscillator signals of the same frequency. This yields quadrature (Q) and in-phase (I) signals, respectively. The latter is a measure for the amplitude of the heterodyne signal, the former is the phase difference between input signal and local oscillator. This phase difference is used in a control loop to adjust the frequency and phase of the local oscillator signal to the input signal. For this purpose the Q signal is amplified by a proportional-integral amplifier, a constant reference frequency is added and the sum is input to the phase increment register (PIR). The PIR, the phase accumulator (PA), and the lookup table for sine/cosine form a numerically controlled oscillator (NCO). The main outputs of the PMS are the values of the PIR and the PA registers, that correspond to frequency and phase of the input signal, respectively. A prototype PMS operating as described above has shown sufficient performance for LISA optical bench testing when operated with electrically generated signals. For optical bench testing sixteen analog input channels will be processed by the FPGA. IV.Test Mass and Telescope SimulatorThe telescope simulator receives a fraction of TX light emitted from the optical bench and interferes it with light that is sent to the science interferometer in the so-called transponder interferometer. The latter acts as reference for the science interferometer. In addition to the transponder interferometer, the telescope simulator houses the so-called commissioning interferometer. It can be used to verify the pathlength noise performance of the telescope simulator itself. A. Purpose and requirementsThe test mass simulator should allow optical pathlength noise measurements to verify The purpose of the telescope simulator is to
We allow for a 10% increase of the requirements by test mass simulator and telescope simulator, respectively. B. Design considerationsTilt actuators are required in the telescope simulator for two purposes:
For both purposes, four degrees of freedom are required each. This is achieved by using pairs of mirrors on tip-tilt mounts, of which the first one is used to set the reflection point on the second one, which in turn sets the direction of the beam. In the test mass simulator for the LISA technology package (LTP) it was observed that PZT stacks only fulfill the LTP length noise requirement for low actuation voltages not sufficient for the complete angle range. Hence, we restricted the choice to those commercially available actuators that do not need a voltage applied to them to stay at an arbitrary angle (set-and-forget devices). Here, the AG-M100N-V6 by Newport was chosen as baseline actuator. The requirements on the RX beam generated by the telescope simulator restrict its power to about 300pW being detected in the science interferometer. Such small light powers are not detectable by IR-viewer cards commonly used to trace the beam path of IR laser beams. The photo diodes to be used in the science interferometer also exhibit a dark current higher than the current that would be generated by a beam of this light power, also rendering them useless in alignment of this beam. These properties of the beam and possible detectors show that no signal can be produced to guide the alignment. The only remaining alternative for alignment is to scan the beam across all possible angles and positions and wait for a “by chance” alignment of the two interfering beams within 500 μrad to allow for efficient heterodyne detection and the generation of differential wavefront sensing (DWS) alignment signals [9], [10]. Obviously, this would be a lengthy and cumbersome procedure which calls for a better way of aligning the two beams. One way to achieve a better alignment is to increase beam power during the alignment stage and then reduce light power afterwards with as little change of geometry and alignment of the RX beam as possible. This allows to detect the beam using conventional methods (IR viewer card and DC-current of photodiode) during the alignment phase and to switch over to heterodyne detection and the use of DWS signals enabled through this pre-alignment. Further information on design and construction of test mass and telescope simulator can be found in [11]. C. LayoutFigure 8 shows the layout of the test mass and telescope simulator. It defines the positions of all optics to be bonded on the Zerodur base plate using hydroxide-catalysis bonding [12] to provide the necessary stability. Due to the adjustable optics the tolerances on these positions are sufficiently loose to enable the use of a template for the positioning during the bonding process which greatly eases and speeds up the building process. The test mass simulator consists of optical components TMSim1 and Mtm. A pick-up mirror will be placed on the LISA OB between PBS2 and the out-of-plane mirror to send light to mirror TMSim1 that reflects the light to mirror Ttm. The latter uses a reflective gold coating and is placed on a tilt actuator and linear actuator to simulate the test mass. This actuator assembly is described in more detail in Sec. IV-D. All other components on the Zerodur baseplate belong to the telescope simulator, that is described in the following. The mainly s-polarised output beam from the Fibre Injector Optical Subassembly (FIOS) Telsim in the upper right corner is reflected from polarising beam splitter PBS101 to clean the polarisation. Beam splitter BS101 splits the beam into two equal parts. One part is magnified by lens1 and lens 2 to 5 mm diameter and reflected by mirrors M4 and M3 (45° angle of incidence). Beam combiner BS6 reflects 1% of the power towards the out-of-plane mirror Mup that sends the RX beam towards the science interferometer on the optical bench. M4 and M3 are mounted in actuator assemblies described in Section IV-D below. They are used to align the RX beam to the science interferometer. The other part of the beam is reflected by mirrors M101 and M102 (41.5° angle of incidence) and directed to beam combiner BS103. There, the beam from the Telsim FIOS interferes with a fraction of the TX beam from the optical bench. One percent of the p-polarised TX beam is reflected by BS6, reduced in diameter by lens2 and lens1, transmitted through PBS102 and rotated to s polarisation by half-wave plate HWP1. The interference signals of TX beam and RX beam are detected by photo detectors PD1 and PD2. Mirrors M101 and M102 are used to align the transponder interferometer. A second FIOS, FIOS com, has been placed in the central upper part. It is used for commissioning of the telescope simulator. Its s-polarised output beam is split by BS104 into two parts. The reflected part propagates towards BS6. For commissioning a mirror and quarter-wave plate will be placed between M3 and BS6, that reflect all light back and rotate its polarisation by 90°, such that it is transmitted through PBS102 and interferes at BS103 with light from FIOS Telsim. The part transmitted through BS104 is interfered at BS106 with light from FIOS Telsim. The interference signals in this commissioning interferometer are detected by photo detectors PD3 and PD4. Polarisation cleaning for the light from FIOS com is performed by PBS102 and PBS105. For commissioning, the beam path between BS102 and BS104 has to be blocked. This will be achieved by insertion of a mirror that reflects the light coming from FIOS Telsim to the space between beam dumps BD1 and BD2 where an additional beam dump will be placed. A fraction of the power emitted from FIOS Telsim and FIOS com is detected by photo detectors behind BS107 and BS105, respectively, to monitor and or stabilise the output power from the FIOSes. D. Actuator assembliesThe optical paths on telescope and test mass simulator must be stable to The actuator assemblies for mirrors M101, M102, M3, and M4 use a metal adapter instead of the linear actuator shown in Fig. 9. E. Performance predictionIn this section we predict the pathlength noise performance of the telescope simulator in combination with the optical bench. We also provide the data required to predict the performance in commissioning mode and of optical bench and telescope simulator individually. 1) Local noise sources: First, we consider noise sources that are local to an interferometric readout: shot noise, electronic noise, and relative power noise. They add to the signal at the heterodyne frequency and degrade the interferometric pathlength measurement. Given a signal and a noise of an interferometer readout, its pathlength noise Δs is given by where λ is the laser wavelength. We consider the signal and the noise from both ports of the interferometer, i.e. from the combined signals of two quadrant photodiodes with four segments each. The rms signal of the combined photocurrents (in A) is given by where η is the efficiency of the photodiodes, γ the mode overlap (heterodyne efficiency) of the interfering beams and Psig and Plo are the signal and local oscillator power before the recombination beamsplitter, respectively. The shot noise, electronic noise, and relative power noise of the combined photocurrents in where qe is the elementary charge, nel is the electronic noise per QPD segment, and nRPN is the relative power noise of signal and local oscillator beam. We use Inserting (3) and (4), (5), (6) into (2) yields the pathlength noise due to shot noise, electronic noise and relative power noise for interferometer I. These three noise sources are independent and hence add quadratically. Their combined pathlength noise ΔsL(I) is given by The optical powers of signal beam and local oscillator beam are listed in Table I. Tables II and III summarise the parameters and constants used in this and the following sections. TABLE ITotal Optical Powers Before Recombination Beamsplitters
TABLE IIQuantities Used for Performance Prediction.
TABLE IIIConstants and Parameters Used for Performance Prediction; Coefficient of Thermal Expansion CTE, Fused Silica FS
We include a further source of pathlength noise that is caused by the point ahead angle mechanism (PAAM) [2] on the optical bench. Optical pathlength noise generated by the PAAM is required to be below ΔsPAAM. This requirement has been verified experimentally [2]. 2) Temperature-related local noise sources: The selected actuators show a coupling from temperature to angle. When reflection point and rotation point do not coincide, the distance l between the two (in a single actuator axis) transforms temperature driven tilts into optical pathlength changes. This pathlength noise Δsrot for a single actuator axis is given by where ϕ is the angle of incidence and Optical pathlength noise due to thermal expansion of the actuator assemblies ΔsAct is given by where 3)Noise sources coupled to two interferometers: Some noise sources couple via a path imbalance in the length measurement. Interferometric testing of the LISA optical bench requires to read out two interferometers simultaneously. In this case, the coupling factor is given by a length difference between the two interferometers. In this section we discuss the relevant length differences and allocate them to either optical bench or telescope simulator. We quantify the influence of laser frequency noise, thermal expansion of the setup (thermoelastic noise) and temperature-driven pathlength changes in fused silica (thermo-optic noise) on the length measurement. Optical pathlength noise due to thermo-optical noise within fused silica ΔsFS is given by where ΔT is the temperature noise, OPDFS the effective imbalance in fused silica, αFS the coefficient of thermal expansion of fused silica, nFS its refractive index, and Optical pathlength noise due to thermo-elastic noise of Zerodur ΔsZ is given by where OPDZ and αZ are the path imbalance and coefficient of thermal expansion of Zerodur. Optical pathlength noise due to laser frequency noise Δsfreq is given by where c is the speed of light and Δf is the laser frequency noise. Figure 10 shows a schematic view on optical bench and telescope simulator. The TX laser is input to the optical bench, the RX laser is input to the telescope simulator. On the optical bench the science interferometer is shown (BS17), on the telescope simulator the transponder interferometer is shown (BS103). We define path imbalances ΔLS and ΔLT between TX and RX laser in science and transponder interferometer, respectively. Table IV lists the endpoints of LTS, LRS, LTT and LRT and all other lengths to be used in this section. TABLE IVPathlength Imbalances and Their Endpoints; CF. Fig. 10
We do not consider the lengths between TX laser and BS13 and RX laser and BS101 because they are common to science and transponder interferometer and hence cancel in the difference of the two. For measurements utilising science and transponder interferometer, the effective length imbalance ΔL is given by By inserting (15), (16), (18) and (19) into (17) we find with We have partitioned the effective path imbalance ΔL into a part on the optical bench (ΔLOB) and a part on the telescope simulator (ΔLTS). This scheme was used to produce the effective path imbalances on Zerodur and in fused silica as listed in Table V. TABLE VEffective Path Mismatch on Zerodur and in Fused Silica; Optical Bench OB, Telescope Simulator TS, Transponder Mode Transp., Commissioning Mode Comm.
4) Combinations of noise sources: In this section we combine the individual noise sources discussed in the previous sections. We assume, all actuator assemblies are identical and exhibit the same temperature noise. Then their pathlength noises add linearly. In the horizontal axis the pathlength noises of M101 and M102 and M3 and M4 cancel each other. In the vertical axis the noises add up: The factor of two outside the round brackets in (23) takes two actuators each into account. The same is true for the combined effect of longitudinal actuator noise with temperature on the optical pathlength noise ΔsAct,all: Now we combine the noise sources discussed previously. Optical pathlength noise associated with the telescope simulator ΔsTS is given by Depending on what path imbalance data from Table V is used, both the pathlength noise performance of the telescope simulator in transponder mode or in commissioning mode can be predicted. The noise attributed to the optical bench ΔsOB is given by Finally, we combine the noise sources of optical bench and telescope simulator and obtain for the noise of optical bench and the telescope simulator in transponder mode ΔsOB+TS. Figure 11 shows this predicted pathlength noise performance over temperature noise. For temperature noise below Fig. 11.Performance prediction for optical bench and telescope simulator as function of temperature noise. ![]() 5) Frequency dependencies: In Fig. 11 we have considered pathlength noise as function of temperature noise. Now we fix the temperature noise to a specific value and look at the frequeny dependencies of the noise contributions. Pathlength noise due to electronic noise, relative power noise and shot noise has a uniform frequency distribution. We can use (8), (9), (7), and (10) for a fixed temperature noise without modification. For the local temperature-dependent noise sources tilt coupling Δsrot and thermal expansion ΔsAct of actuator assemblies we define an additional noise shape uM(f) that we multiply with (23) and (24). We use a temperature noise of For the pathlength noise due to the PAAM, ΔsPAAM, and frequency noise induced pathlength noise, Δsfreq, we multiply (11) and (14) with the shape factor uPL(f). For pathlength noise on Zerodur, ΔsZ, and pathlength noise in fused silica, ΔsFS, (13) and (12) are multiplied with uM(f). Figure 12 shows the resulting predicted pathlength noise as function of frequency for V.ConclusionTesting of the LISA optical bench elegant breadboard requires to externally complete the test mass and science interferometers by simulators. We have presented such simulators that will be implemented on a common Zerodur baseplate. The test mass simulator consists of a gold-coated mirror on a tilt and linear actuator. The telescope simulator works as a transponder and provides a reference for the science interferometer on the LISA OB. Both simulators employ tilt actuators for beam alignment that are held in thermally compensating mounts. The pathlength noise performance of the telescope simulator can be verified independently from the LISA OB by means of a dedicated interferometer on the simulator. In the experiments, temperature noise control will be crucial. On the basis of the measured temperature noise and the performance predictions presented above characterisation of the LISA optical bench to its requirements seems possible. A tracking phasemeter implemented in a field-programmable gate array (FPGA) will be used as phase measurement system. Prototypes of this phasemeter have shown the required phase noise performance using electrical signals. A 16 channel version of the phasemeter will be used, sufficient to read out two fully redundant interferometers equipped with quadrant photo detectors. AcknowledgmentWe acknowledge funding by the European Space Agency within the project “Optical Bench Development for LISA”, support from STFC and UKSA, and support by Deutsches Zentrum für Luft und Raumfahrt (DLR) with funding from the Bundesministerium für Wirtschaft und Technologie (DLR project reference 50 OQ 0601). We thank the German Research Foundation for funding the cluster of Excellence QUEST - Centre for Quantum Engineering and Space-Time Research. ReferencesO. Jennrich,
“LISA technology and instrumentation,”
Class. Quantum Grav., 26
(15),
(2009). https://doi.org/10.1088/0264-9381/26/15/153001 Google Scholar
N. R. J.A.C.M. Pijnenburg,
“Picometer stable scan mechanism for gravitational wave detection in space: LISA PAAM,”
in Proc. of the International Conference on Space Optics,
(2010). Google Scholar
L. d’Arcio, J. Bogenstahl, M. Dehne, C. Diekmann, E. D. Fitzsimons, R. Fleddermann, E. Granova, G. Heinzel, H. Hogenhuis, C. J. Killow, M. Perreur-Lloyd, J. Pijnenburg, D. I. Robertson, A. Shoda, A. Sohmer, A. Taylor, M. Tröbs, G. Wanner, H. Ward, and D. Weise,
“OPTICAL BENCH DEVELOPMENT FOR LISA,”
in Proc. of the International Conference on Space Optics,
(2010). Google Scholar
“LISA Assessment Study Report,”
ESA document ESA/SRE,
(3),
(2011). Google Scholar
“NGO Assessment Study Report,”
ESA document ESA/SRE,
(19),
(2012). Google Scholar
J. H. Shirley,
“Modulation transfer processes in optical heterodyne saturation spectroscopy,”
Opt. Lett., 7
(11), 537
–539
(1982) http://ol.osa.org/abstract.cfm?URI=ol-7-11-537 Google Scholar
G. Heinzel, V. Wand, A. Garcia, O. Jennrich, C. Braxmaier, D. Robertson, K. Middleton, D. Hoyland, A. Rüdiger, R. Schilling, U. Johann, and K. Danzmann,
“The LTP interferometer and phasemeter,”
Class. Quantum Grav., 21 S581
–S587
(2004). https://doi.org/10.1088/0264-9381/21/5/029 Google Scholar
I. Bykov, J. J. E. Delgado, A. F. G. Marin, G. Heinzel, and K. Danzmann,
“LISA phasemeter development: Advanced prototyping,”
in Journal of Physics: Conference Series,
012017
(2009). Google Scholar
E. Morrison, B. J. Meers, D. I. Robertson, and H. Ward,
“Automatic alignment of optical interferometers,”
Appl. Opt., 33 5041
–5049
(1994). https://doi.org/10.1364/AO.33.005041 Google Scholar
E. Morrison, B. J. Meers, D. I. Robertson, and H. Ward,
“Experimental demonstration of an automatic alignment system for optical interferometers,”
Appl. Opt., 33 5037
–5040
(1994). https://doi.org/10.1364/AO.33.005037 Google Scholar
J. Bogenstahl, M. Tröbs, L. dArcio, C. Diekmann, E. D. Fitzsimons, J.-S. Hennig, F. G. Hey, C. J. Killow, M. Lieser, S. Lucarelli, M. Perreur-Lloyd, J. Pijnenburg, D. I. Robertson, A. Taylor, H. Ward, D. Weise, G. Heinzel, and K. Danzmann,
“Design and construction of a telescope simulator for LISA optical bench testing,”
in Proc. of the International Conference on Space Optics,
(2012). Google Scholar
E. J. Elliffe, J. Bogenstahl, A. Deshpande, J. Hough, C. Killow, S. Reid, D. Robertson, S. Rowan, H. Ward, and G. Cagnoli,
“Hydroxide-catalysis bonding for stable optical systems for space,”
Classical and Quantum Gravity, 22
(10), S257
(2005) http://stacks.iop.org/0264-9381/22/i=10/a=018 Google Scholar
F. Guzmán Cervantes,
“Gravitational wave observation from space: Optical measurement techniques for LISA and LISA Pathfinder,”
Ph.D. dissertation, Leibniz Universität Hannover
(2009). Google Scholar
|