28 May 2019 Analysis of scatter artifacts in cone-beam CT due to scattered radiation of metallic objects
Domenico Iuso M.D., Robert Frysch M.D., Tim Pfeiffer, Georg Rose
Author Affiliations +
Proceedings Volume 11072, 15th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine; 110721K (2019) https://doi.org/10.1117/12.2534465
Event: Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 2019, Philadelphia, United States
Abstract
Cone-beam computed tomography (CBCT) is a widely used technique for diagnostic or monitoring purposes. Compared to the traditional CT, a CBCT is more affected by scatter artifacts because of the large volume being irradiated by the beam. The research divulged in this paper is about the assessment of the influence that metallic implants may have on degrading image quality of CBCT due to scattered radiation. The evaluation method is based on Monte-Carlo (MC) simulations of the physical processes that X-ray photons undergo in typical CBCT setups, in presence and absence of highly scattering metallic implants (coils used for treatment of aneurysms and pacemakers). The results show that the scattered radiation caused by metallic objects and reaching the detector produces slight degradation of CBCT image quality and, moreover, it is demonstrated that the intrinsic absorption and beam-hardening effect of these implants have bigger impact on the overall image fidelity.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Domenico Iuso M.D., Robert Frysch M.D., Tim Pfeiffer, and Georg Rose "Analysis of scatter artifacts in cone-beam CT due to scattered radiation of metallic objects", Proc. SPIE 11072, 15th International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 110721K (28 May 2019); https://doi.org/10.1117/12.2534465
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Photons

Head

Sensors

Scattering

Image quality

Monte Carlo methods

X-ray computed tomography

RELATED CONTENT

Dual layer x-ray detector simulation
Proceedings of SPIE (March 16 2020)
The new device for imaging: Volumescope
Proceedings of SPIE (December 06 2006)
Effects of scattered x-rays on cone-beam CT images
Proceedings of SPIE (May 01 1994)
Spectral analysis of scattered radiation in CT
Proceedings of SPIE (March 18 2008)

Back to Top