Near infrared light generated by specialized instrumentation was passed through artificially oxygenated human blood during simultaneous sampling by a co-oximeter. Characteristic absorption spectra were analyzed to calculate the ratio of oxygenated to reduced hemoglobin. A positive linear regression fit between diffuse transmission oximetry and measured blood oxygenation over the range 23% to 99% (r2 equals .98, p < .001) was noted. The same technology was used to pass two channels of light through the scalp of brain-injured patients with prolonged, decreased level of consciousness in a tertiary care neuroscience ICU. Transmission data were collected with gross superficial-to-deep spatial resolution. Saturation calculation based on the deep signal was observed in the patient over time. The procedure was able to be performed clinically without difficulty; rSO2 values recorded continuously demonstrate the usefulness of the technique. Using the same instrumentation, arterial input and cerebral response functions, generated by IV tracer bolus, were deconvoluted to measure mean cerebral transit time. Date collected over time provided a sensitive index of changes in cerebral blood flow as a result of therapeutic maneuvers.
|