Paper
14 October 1994 Optimum microchannel plate (MCP) configuration for use in high-speed, high-resolution x-ray imaging
Steve E. Grantham, Edward J. Miesak, Paul C. T. Reese, Martin C. Richardson
Author Affiliations +
Abstract
Microchannel plates (MCPs) are incorporated in a wide variety of x-ray imaging and detection devices. Experimental measurements are presented in this paper which are used to determine the MCP and phosphor configuration for optimizing spatial resolution, temporal resolution, and gain in an x-ray framing camera. We investigate with the use of pulsed electron accelerating voltages to attain extraction fields higher than those possible at safe DC levels. In addition, MCPs with exit faces coated with an insulating material, in order to increase the maximum safe DC electron extraction field, were also tested.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Steve E. Grantham, Edward J. Miesak, Paul C. T. Reese, and Martin C. Richardson "Optimum microchannel plate (MCP) configuration for use in high-speed, high-resolution x-ray imaging", Proc. SPIE 2273, Ultrahigh- and High-Speed Photography, Videography, and Photonics '94, (14 October 1994); https://doi.org/10.1117/12.189019
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Microchannel plates

Cameras

Capillaries

X-ray imaging

Coating

Spatial resolution

X-rays

RELATED CONTENT

X-Ray Gated Images Of Imploding Microballoons
Proceedings of SPIE (February 01 1989)
A large-format gated x-ray framing camera
Proceedings of SPIE (January 28 2004)
X-Ray Framing Camera Images Of Imploding Targets
Proceedings of SPIE (November 27 1989)
Hybrid scintillators for x-ray imaging
Proceedings of SPIE (April 11 1996)

Back to Top