Paper
16 July 1981 Digital Radiography Using The Quantex DS-20
G. Allan Johnson, K. Ford, R. Heinz
Author Affiliations +
Proceedings Volume 0273, Application of Optical Instrumentation in Medicine IX; (1981) https://doi.org/10.1117/12.931792
Event: Application of Optical Instrumentation in Medicine, 1981, San Francisco, United States
Abstract
The Quantex DS-20 digital image processor has been evaluated as a low cost alternative for digital vascular radiography. The processor is used in conjunction with a conventional remote radiographic/fluoroscopic system with no generator/processor interface required. (Use of a video-tape system reduces timing problems with minimal image degradation). Phantom studies indicate vessels as small as 1.5 mm can be seen with concentrations of contrast as low as 10 mgΓ/ml. Animal studies show that subtraction images of diagnostic quality can be generated.
© (1981) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G. Allan Johnson, K. Ford, and R. Heinz "Digital Radiography Using The Quantex DS-20", Proc. SPIE 0273, Application of Optical Instrumentation in Medicine IX, (16 July 1981); https://doi.org/10.1117/12.931792