19 July 1996 Transmission maps of the ACIS UV/optical blocking filters
Leisa K. Townsley, Forbes R. Powell, James F. MacKay, Max G. Lagally, John A. Nousek, Gordon P. Garmire
Author Affiliations +
Abstract
The AXAF CCD imaging spectrometer (ACIS) consists of ten CCDs arranged in two arrays, one for imaging astronomical fields and one to be used in conjunction with transmission gratings for spectroscopy of astrophysical sources. ACIS uses Lexan/aluminum meshless films placed above these two CCD arrays to filter by mapping their soft x-ray transmission on fine spatial scales, so that the filter response can be removed from the CCD data and a more accurate estimate of the true sky recovered. We measured engineering and flight versions of these filters at the University of Wisconsin Synchrotron Radiation Center between June 1995 and April 1996. For all data, better than one percent accuracy in transmission as a function of energy was maintained over the entire filter area. The resulting transmission maps reveal spatial non-uniformities in the filters of about 0.5 percent to 2 percent. These transmission maps provide the finest spatial calibration ever achieved on such filters.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Leisa K. Townsley, Forbes R. Powell, James F. MacKay, Max G. Lagally, John A. Nousek, and Gordon P. Garmire "Transmission maps of the ACIS UV/optical blocking filters", Proc. SPIE 2805, Multilayer and Grazing Incidence X-Ray/EUV Optics III, (19 July 1996); https://doi.org/10.1117/12.245086
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical filters

Image filtering

Spectroscopy

Imaging arrays

Charge-coupled devices

Aluminum

Imaging spectroscopy

RELATED CONTENT

Xtend, the soft x ray imaging telescope for the X...
Proceedings of SPIE (August 31 2022)
ASTRO-E/XRS blocking-filter calibration
Proceedings of SPIE (October 22 1999)
Soft x-ray and EUV efficiencies of CCDs
Proceedings of SPIE (November 19 1993)
The AXAF CCD Imaging Spectrometer
Proceedings of SPIE (July 14 1986)

Back to Top