2 May 1997 Performance of a CsI:TL-screen-based scanning slot detector for digital mammography
Zhenxue Jing, Walter Huda, James K. Walker, Won Y. Choi
Author Affiliations +
Abstract
Thallium activated CsI scintillation screen (CsI:Tl) has advantages over rare-earth phosphor screens when used in a scanning slot x-ray detector for digital mammography. The scintillation decay time for CsI:Tl is only approximately 1 microsecond(s) which eliminates the afterglow effect associated with the use of rare-earth phosphor. The CsI needles serve to limit the spread of the scintillation light which permits the use of a relatively thick CsI:Tl screen to improve detector x-ray interaction efficiency without sacrificing resolution performance. A prototype scanning slot detector was made of a CsI:Tl screen and was optically coupled to two CCDs by plastic optical fiber image guides. The screen was composed of prismatic CsI crystals with a needle size of approximately 5 micrometers . Image guides used in the detector had an input to output ratio of 1:1, and were made of 10 micrometers diameter plastic optical fibers. Each CCD, operated in the time-delayed integration mode, was an 1100 X 330 pixel array with pixel size of 24 micrometers . A full scale scanning slot detector will contain eight rather than two such modules. The x-ray interaction efficiency of the slot x-ray detector was calculated to be approximately 84 percent at 20 keV for the CsI:Tl screen, and was approximately 20 percent greater than that of a 31.7 mg/cm2 thick Gd2O2S:Tb phosphor screen. Limiting spatial resolution was investigated by taking the images of a 1 degree star resolution test pattern as a function of detector scan speed. Limiting spatial resolution was approximately 15 1p/mm at a scan speed of 4 cm/s and improved only slightly to approximately 16 1p/mm as the scanning speed decreased to 1 cm/s.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhenxue Jing, Walter Huda, James K. Walker, and Won Y. Choi "Performance of a CsI:TL-screen-based scanning slot detector for digital mammography", Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.273994
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

X-ray detectors

Sensors

Charge-coupled devices

Scintillation

Spatial resolution

Digital mammography

RELATED CONTENT

The use of EM CCDs on high resolution soft x...
Proceedings of SPIE (September 13 2011)
The use of CCDs and EM CCDs for future soft...
Proceedings of SPIE (September 17 2012)
Digital Slot Scan Mammography Using CCDs
Proceedings of SPIE (January 01 1987)

Back to Top