6 October 1999 Development of the photon-counting 256ch CdTe line detector
Kenji Tokumori, Fukai Toyofuku, Shigenobu Kanda, Sueki Baba, Yoshio Mito, Kazuyuki Hyodo, Masami Ando, Chikao Uyama
Author Affiliations +
Abstract
We have developed a photon-counting 256ch CdTe line detector system for a monochromatic x-ray CT system using fluorescent x-rays generated by synchrotron radiation. The size of each detector element is 1.98 mm(w) X 1.98 mm(h) X 0.5 mm(t). Each element has two discriminators (an upper and a lower discriminator) and two 16-bit counters (an upper and a lower counter). Each discriminator rejects pulses having a pulse height lower than the chosen voltage limits. All pulses in between the upper and lower voltage limits were obtained by subtracting the upper counter value from the lower counter value. By changing the voltage limits, we can obtain an incident x-ray energy spectrum. Several energy spectra for the fluorescent x-ray and standard (gamma) -ray sources were measured by using this detector. The detector showed a sufficient energy resolution, and has been found to be suitable as a detector of monochromatic x-ray CT.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kenji Tokumori, Fukai Toyofuku, Shigenobu Kanda, Sueki Baba, Yoshio Mito, Kazuyuki Hyodo, Masami Ando, and Chikao Uyama "Development of the photon-counting 256ch CdTe line detector", Proc. SPIE 3770, Medical Applications of Penetrating Radiation, (6 October 1999); https://doi.org/10.1117/12.368180
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-rays

X-ray detectors

Synchrotron radiation

X-ray sources

Calibration

X-ray computed tomography

RELATED CONTENT

3D x-ray reconstruction using lightfield imaging
Proceedings of SPIE (September 05 2014)
Ground Calibration of the IXPE Optics and Telescope
Proceedings of SPIE (January 01 1900)
Fan beam monochromatic x ray CT using fluorescent x rays...
Proceedings of SPIE (October 06 1999)
Design and evaluation of a prototype volume CT scanner
Proceedings of SPIE (April 20 2005)

Back to Top