5 January 2001 Zone plate achromatic doublets
Alan G. Michette, Christopher J. Buckley, Federico Gallo, Keith Powell, Slawka J. Pfauntsch
Author Affiliations +
Abstract
Zone plates as used in high resolution x-ray microscopy have focal lengths which are proportional to the x-ray energy. This means that in techniques such as elemental and chemical state mapping, which require images to be made at more than one energy, refocusing is required which can lead to loss of image registration. Using two zone plates, in a similar fashion to achromatic refractive lens doublets for visible light, it is possible to focus two x-ray energies to the same spot.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Alan G. Michette, Christopher J. Buckley, Federico Gallo, Keith Powell, and Slawka J. Pfauntsch "Zone plate achromatic doublets", Proc. SPIE 4145, Advances in X-Ray Optics, (5 January 2001); https://doi.org/10.1117/12.411651