15 April 2008 A quantitative study of the practical sensitivity limit of a terahertz absorption spectrometer
Author Affiliations +
Abstract
In gas spectroscopy, chemicals can be identified by the set of frequencies at which their absorption lines occur. The concentration can be quantitatively estimated from the intensity of any of the absorption lines. The sensitivity of the spectrometer, i.e., the minimum detectable concentration, is ideally limited by the ratio of the source power to detector noise-equivalent power. In practice, the sensitivity is usually orders of magnitude worse due to systematic effects. In this work we built a simple gas terahertz transmission spectrometer to analyze how the source output power stability, the detector sensitivity, and atmospheric pressure affect its sensitivity. As a test gas we used methyl chloride in a mixture with air and modifid the widths of the absorption lines by changing partial pressure of air. This demonstration of a simple absorption spectrometer gives us insight into the approach to making a highly sensitive terahertz spectrometer.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jon E. Bjarnason, Charles Dietlein, and Erich N. Grossman "A quantitative study of the practical sensitivity limit of a terahertz absorption spectrometer", Proc. SPIE 6949, Terahertz for Military and Security Applications VI, 694908 (15 April 2008); https://doi.org/10.1117/12.784620
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Protactinium

Absorption

Spectroscopy

Sensors

Neptunium

Amplifiers

Signal attenuation

Back to Top