20 November 2009 Experimental study on absolute spectral responsivity value transfer with cryogenic radiometer
Xueshun Shi, Kunfeng Chen, Hongyuan Liu, Hengfei Wang, Quanshe Sun
Author Affiliations +
Abstract
The spectral responsivity measurement theoretical outline was demonstrated. Absolute spectral responsivity of the Silicon trap detector is calibrated with laser intensity stabilizer, closed loop refrigeration cryogenic radiometer. Seven wavelength laser beams of 476.1nm, 488nm, 514.7nm, 568nm, 632.8nm, 647.1nm and 1064nm, are chosen for the calibration experiments. Relative spectral responsivity of the cavity pyroelectric detector is calibrated in the band from 600nm to 15μm. Absolute spectral responsivity of the pyroelectric detector is transferred at 1064nm. The absolute and relative measurement uncertainty is evaluated. Results demonstrated that the absolute spectral responsivity of the Silicon trap detector and the cavity pyroelectric detector is 0.3954A/W, whose uncertainty is better than 0.04%, and 5.4E-7A/W, whose uncertainty is better than0.4%, respectively, at 1064nm. The pyroelectric detector absolute spectral responsivity uncertainty is better than 5% in other wavelength.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Xueshun Shi, Kunfeng Chen, Hongyuan Liu, Hengfei Wang, and Quanshe Sun "Experimental study on absolute spectral responsivity value transfer with cryogenic radiometer", Proc. SPIE 7511, 2009 International Conference on Optical Instruments and Technology: Optoelectronic Measurement Technology and Systems, 75111A (20 November 2009); https://doi.org/10.1117/12.837791
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Cryogenics

Radiometry

Pyroelectric detectors

Calibration

Silicon

Laser stabilization

Back to Top