Paper
17 September 2012 Vacuum facility for calibration of space instrumentation in cleanroom
Author Affiliations +
Abstract
Calibration of optical systems is a fundamental step in the development of a space instrumentation. We have built a new cleanroom environment, divided in different areas characterized by a different level of contamination control. A vacuum chamber (a tube of 80 cm diameter, and 2 m length), able to accommodate optical components as well as whole instruments, is interfaced with a ISO6 area, allowing the handling of the instrumentation in a clean environment. The vacuum system is dimensioned to reach 10-7 mbar pressure in the chamber. Inside, a two axis platform allows the rotation of the instrument with respect to the incident collimating beam, in order to test the response of the instrument to light coming from different points of the field of view. A monochromator coupled with different sources provides radiation in the 40-350 nm spectral range, while a parabolic mirror is used as a collimator. As source, different spectral lamps can be used to generate emission lines, while a Xe lamp can be used to have continuum spectrum. An high brilliant hollow cathode lamp has been fabricated by the group to generate extreme ultraviolet radiation. Different calibrated detectors and other completing optical components are available.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. G. Pelizzo, P. Zuppella, V. Polito, A. J. Corso, S. Zuccon, and P. Nicolosi "Vacuum facility for calibration of space instrumentation in cleanroom", Proc. SPIE 8443, Space Telescopes and Instrumentation 2012: Ultraviolet to Gamma Ray, 84432Q (17 September 2012); https://doi.org/10.1117/12.927294
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Lamps

Monochromators

Aerospace engineering

Optical components

Extreme ultraviolet

Spectroscopy

Back to Top