19 March 2013 Optimized control of a dynamic, prepatient attenuator
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86681Q (2013) https://doi.org/10.1117/12.2008229
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Dynamic attenuators are beam shaping filters that can customize the x-ray illumination field to the clinical task and for each view. These dynamic attenuators replace traditional attenuators (or “bowtie filters”) and decrease radiation dose, dynamic range, and scatter when compared to their static counterparts. We propose a one-dimensional dynamic attenuator that comprises multiple wedges with axially-dependent triangular cross-sections, and which are translated in the axial direction. These wedges together produce a time-varying, piecewise-linear attenuation function. We investigate different control methods for this attenuator and estimate the ability of the dynamic attenuator to reduce dose while maintaining the peak variance of the scan. With knowledge of the patient anatomy, the dynamic attenuator can be controlled by solving a convex optimization problem. This knowledge could be determined from a low dose pre-scan. Absent this information, various heuristics can be used. We simulate the dynamic attenuator on datasets of the thorax, abdomen, and a targeted scan of an abdominal aortic aneurysm. The dose and scatter-to-primary ratio (SPR) are estimated using Monte Carlo simulations, and the noise is calculated analytically. Compared to a system using the standard bowtie with typical mA modulation, dose reductions of 50% are observed. Compared to an optimized, patientspecific mA modulation, the typical dose reduction is 30%. If the dynamic attenuator is controlled with a heuristic, typical dose reductions are also 30%. The gains are larger in the targeted scan. The SPR is also reduced by 20% in the abdomen. We conclude that the dynamic attenuator has significant potential to reduce dose without increasing the peak variance of the scan.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Scott S. Hsieh and Norbert J. Pelc "Optimized control of a dynamic, prepatient attenuator", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86681Q (19 March 2013); https://doi.org/10.1117/12.2008229
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Attenuators

Modulation

Signal attenuation

Image quality standards

Monte Carlo methods

Abdomen

Convex optimization

RELATED CONTENT

Fluid-filled dynamic bowtie filter: a feasibility study
Proceedings of SPIE (March 18 2015)
Algorithms for optimizing CT fluence control
Proceedings of SPIE (March 19 2014)
Implementation of a piecewise-linear dynamic attenuator
Proceedings of SPIE (March 09 2018)
A new concept for fluence field modulation in x ray...
Proceedings of SPIE (March 16 2020)

Back to Top