4 March 2014 Spectrum correction considering light source fluctuation for non-invasive blood glucose sensing
Author Affiliations +
Abstract
The purpose of this study is to correct baseline shift in absorbance spectrums caused by light source fluctuation. To improve quantitative evaluation performance of blood glucose level, baseline shift is corrected by multiple scatter correction (MSC). Moreover, to increase the effect of the MSC, water vapor absorbance is subtracted, and relative glucose absorbance are calculated by dividing with hemoglobin absorbance at 1544 [cm-1]. In order to verify the effectiveness of the proposed spectrum correction method, light source fluctuation is simulated on the Fourier transform infrared spectroscopy (FT-IR), and we apply the proposed method to the spectrums measured by FT-IR. From the simulation results, the baseline shift was successfully reduced by proposed method.
© (2014) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Satoru Suzuki, Akane Ishida, Pradeep K. W. Abeygunawardhana, Kenji Wada, Akira Nishiyama, and Ichiro Ishimaru "Spectrum correction considering light source fluctuation for non-invasive blood glucose sensing", Proc. SPIE 8936, Design and Quality for Biomedical Technologies VII, 893604 (4 March 2014); https://doi.org/10.1117/12.2038900