Significance: Spatial frequency domain imaging (SFDI) is a diffuse optical measurement technique that can quantify tissue optical absorption (μa) and reduced scattering ( Aim: We describe the design of openSFDI and report on the accuracy and precision of optical property extractions for three different systems fabricated according to the instructions on the openSFDI website. Approach: Accuracy was assessed by measuring nine tissue-simulating optical phantoms with a physiologically relevant range of μa and Results: The openSFDI systems had an error of 0 ± 6 % in μa and −2 ± 3 % in Conclusion: The openSFDI provides a customiz |
1.IntroductionSpatial frequency domain imaging (SFDI) is a noninvasive, label-free, diffuse optical technique that can generate two-dimensional maps of the absorption coefficient () and reduced scattering coefficient () of biological tissue. To accomplish this, SFDI utilizes sinusoidally modulated illumination patterns, which are projected onto the sample, and the remitted light is detected with a camera. A demodulation and calibration procedure is used to estimate the spatial modulation transfer function (MTF) of the sample. The MTF can then be fit to a computational model of photon transport to determine both and on a pixel-by-pixel basis.1 Knowledge of at several different carefully chosen wavelengths allows for the molar concentration of tissue chromophores such as oxyhemoglobin and deoxyhemoglobin to be estimated. Both optical properties and chromophore concentrations can then be used as optical biomarkers for a wide range of applications. The ability of SFDI to provide quantitative physiological and molecular information in tissue, combined with its relative simplicity, safety, and low cost, has led to its use for a variety of biomedical applications. These include perfusion monitoring of skin flaps during graft surgery,2–4 investigations of blood flow in the brain,5 and assessment of resected tissues following surgery.6,7 The relatively shallow penetration depth of SFDI (typically less than 5 mm)8 makes this technique particularly well suited for assessing skin conditions such as port wine stain birthmarks,9,10 burns,11–13 scarring,14 pressure ulcers,15 nonmelanoma skin cancer,16 and diabetic vascular disease.17 SFDI has been used in the preclinical setting to noninvasively monitor tumor hemodynamics during treatment in vivo.18 SFDI has also been used in nonbiomedical applications, including the assessment of fruit quality in apples19 and peaches.20 For more details regarding the theory and applications of SFDI, we direct readers to two excellent recent reviews.21,22 The goal of this paper and associated website is to provide a complete guide for constructing an SFDI device (shown schematically in Fig. 1), acquiring SFDI data, and determining optical properties in tissue. In the following sections, we describe the key components of the openSFDI system, outline the data-processing pipeline, and present results of the characterization of three openSFDI systems that have recently been constructed. Our hope is that the openSFDI project will reduce the barriers for research groups interested in incorporating SFDI into their research programs. 2.Spatial Frequency Domain Imaging HardwareThere are three main components to an SFDI system: the light source, the method of spatially modulating the illumination field, and the detector. The specific components of the openSFDI system were chosen to balance availability, cost, and overall performance. A rendering of the openSFDI system that includes all the components is shown in Fig. 2. Each openSFDI component will be discussed in the following subsections, followed by an analysis of the overall system performance. A complete materials list is available at Ref. 23. Whenever possible, components were sourced from commonly used suppliers in the United States to make the purchasing process easier. At the time of this writing, the total cost of an openSFDI system was US$4717. 2.1.Illumination SourceThe first published SFDI systems used a broadband mercury lamp as a light source combined with a tunable emission filter for wavelength selection.1 Subsequent systems have commonly utilized laser diodes or an array of light-emitting diodes (LEDs) for illumination, eliminating the need for filtering.24,25 OpenSFDI uses three high-powered near-infrared LEDs with center wavelengths of 660, 735, and 865 nm. These wavelengths were chosen to straddle the isosbestic point of oxyhemoglobin and deoxyhemoglobin at 800 nm to enable high sensitivity to changes in these chromophores. We note that LED center wavelengths can vary significantly, even for LEDs fabricated from a single manufacturer. Accurate knowledge of the illumination wavelength is critical to ensure an accurate extraction of chromophores from the sample. To account for variability in LED wavelength, we recommend that a spectrometer be used to characterize the center wavelength of each LED used in the openSFDI system and that subsequent data-processing steps assume the measured rather than manufacturer specified wavelength. In addition, LED optical output power can be unstable due to thermal effects and/or instability of the driving current. OpenSFDI utilizes heatsinks to stabilize the temperature of the LEDs, but we recommend that the stability of each LED source be tested and characterized for each system (see Sec. 5.2). This step is particularly important if the target application includes longitudinal repeated measurements or long exposure times. Both of these scenarios can lead to LED temperature fluctuations resulting in unstable light output. 2.2.Spatial ModulationSpatially modulated illumination is required to measure the MTF of the sample for SFDI. A simple method to generate illumination patterns is to pass light through a suitably patterned optical transparency.26,27 However, this method lacks the flexibility to rapidly alter the illumination pattern, reducing its utility for many applications. Most published SFDI systems utilize digital micromirror devices (DMDs) to project patterns onto the sample. DMDs have an array of mirrors that can be individually addressed and tilted to direct and pattern the illumination field. The DMD chosen for openSFDI (LC4500, Keynote Photonics) is available as a small DMD chipset with a 0.45" WXGA resolution DMD. The DMD controller is provided on a separate printed circuit board. The software interface provides low-level access to the DMD hardware which is critical to ensure that no automatic corrections are applied to the input image. For example, gamma correction, which is commonly used to improve image quality for human visualization, can result in unwanted harmonics in the projected sinusoidal pattern and lead to demodulation issues. The openSFDI website provides additional examples of distorted projections as well as instructions for troubleshooting these issues.28 2.3.DetectionThe camera selected for openSFDI (BlackFly-S BFS-U3-13Y3M-C, FLIR, Wilsonville, Oregon) is a , 10-bit, monochromatic CMOS camera chosen because it is relatively low cost and can easily interface with LabVIEW. In addition, it allows for low-level control of the image acquisition which makes it possible to turn off automatic corrections that might lead to nonlinear output. We have also constructed openSFDI systems using scientific CMOS (sCMOS) cameras, as discussed in later sections. Crossed linear polarizers are used in the system to limit the effect of specular reflection from the sample. 3.AcquisitionOpenSFDI was designed to make it easy to rapidly begin acquiring SFDI data. To that end, the website provides LabVIEW acquisition code that will run out of the box with any camera supported by IMAQdx. With modifications, the software may also function with cameras supported by the older IMAQ infrastructure. The software has options for collecting multiple measurements at user-defined intervals, selecting wavelengths, choosing spatial frequencies, and allows users to choose the orientation of the spatially projected patterns. The pattern orientation is critical for making profilometric measurements (see Sec. 5.3.2). The openSFDI system will produce an illuminated area of about , if the instructions on the website are followed closely. The working distance between the sample plane and the detector as well as the field of view depends on the choice of lens mounted to the camera. At an aperture of , typical exposure times are on the order of 100 ms per image depending on the sample and wavelength. 4.Data Processing4.1.CalibrationMost diffuse optical techniques, including SFDI, rely on a calibration measurement to determine the instrument response function (IRF). Once known, the IRF can be removed from subsequent measurements, a step needed to obtain accurate estimates of and . Calibration is typically performed using a tissue-mimicking optical phantom with well characterized and at the illumination wavelengths. Data taken on the calibration phantom are compared with the theoretical diffuse reflectance () values associated with the known calibration phantom optical properties to calculate the IRF. A forward model of photon propagation that maps and to in the spatial frequency domain is used to obtain the theoretical calibration values and determine the IRF. Accurate characterization of the calibration phantom is critical to obtaining accurate SFDI results. Optical phantoms can be fabricated by adding an absorbing agent and a scatterer to an optically neutral substrate such as silicone29,30 Time-resolved and frequency-domain diffuse optical measurements, as well as inverse adding doubling, are generally considered accurate methods for determining the optical properties of a calibration phantom.31,32 Some works have utilized intralipid as a calibration standard, as it is subject to a highly controlled manufacturing environment, and both and of standard intralipid emulsions have been characterized in literature.33,34 We note that characterized calibration phantoms may be purchased from the Institut National d’Optique (Biomimic™),35 but we have not tested or utilized these phantoms for this work. New calibration phantoms may also be characterized with existing SFDI devices, provided the system was adequately calibrated using a known standard. Calibration phantoms produced in this way may suffer from the propagation of measurement errors. The calibration phantom used in this study was provided and characterized by Modulim Inc. (Irvine, California). It had dimensions of () and a of and a of at 800 nm. 4.2.Optical Property EstimatesThe procedure for estimating optical properties from SFDI data has been previously described in detail.1 Briefly, sinusoidal illumination patterns are projected onto the sample at three phases (0, , and ). The resulting collected images are then computationally demodulated to isolate the tissue response function at the projected spatial frequency. The IRF is removed by comparison with a calibration phantom (Sec. 4.1) to calculate the as a function of spatial frequency (i.e., the MTF). Measurement of the MTF at two or more spatial frequencies allows the and of the sample to be calculated on a pixel-by-pixel basis. The choice of spatial frequencies for a given application is multifaceted and depends on the optical properties of the sample36 and the partial volume effects caused by using both a low spatial frequency (which is generally more deeply penetrating) and a high spatial frequency (which is generally less deeply penetrating) together for a single SFDI measurement.8 Our lab uses DC and for several of our projects,18 though the use of AC frequencies up to about are also common.37,38 The openSFDI website provides a step-by-step guide to developing an SFDI processing pipeline,39 as well as an executable application for validating processing code written in-house. In addition, the code utilized to generate the figures in this paper is available upon request. Knowledge of at several wavelengths allows the concentration of tissue chromophores to be calculated. Using extinction spectra of oxyhemoglobin and deoxyhemoglobin from literature,40 the concentration of each of these chromophores can be determined by finding the linear combination of concentrations that most closely match the collected data. This calculation is done by solving a system of linear equations. An example of openSFDI images of the back of a hand in vivo is shown in Fig. 3. 5.Characterization5.1.AccuracyThe accuracy of three different openSFDI systems was assessed by comparing them with a commercial SFDI device (Reflect RS, Modulim, Irvine, California). Nine tissue-mimicking optical phantoms containing different amounts of nigrosin and titanium dioxide nanoparticles, to tune absorption and scattering, respectively,41 were measured by each system at spatial frequencies of 0 and . The phantoms were fabricated to have absorption and scattering that spanned a physiologically relevant range ( and ). Measurements were repeated three times for each instrument, and all the raw data were processed identically. Optical properties were estimated from diffuse reflectance values using a two spatial frequency lookup table generated from Monte Carlo simulations.42 The three systems followed the same overall design as the schematic in Fig. 1 but used different cameras. The first system used a FLIR BlackFly-S camera (FLIR, Wilsonville, Oregon), the second system utilized an Andor sCMOS camera (Zyla, Oxford Instruments, Oxford), and the third system used a Thorlabs sCMOS device (Quantalux, Thorlabs, Newton, New Jersey). The first system was constructed at Boston University by S.M.T. and M.B.A., the second system was also fabricated at Boston University by K.K., while the third was initially constructed at the University of Maine by W.A. and K.T. and transported to Boston University for characterization. Interpolation of and values from the commercial instrument was conducted to account for the fact that the wavelengths used for openSFDI (660, 735, and 865 nm) were different from the commercial device (621, 691, 731, 811, and 851 nm). A cubic spline was used to fit the absorption spectrum measured by the commercial system and values at the openSFDI wavelengths were found via interpolation. For , the scattering spectrum measured by the commercial device was fit to the equation , where represents the wavelength, was set at 800 nm, and and were constants. Different and values were found for each phantom. The values for the openSFDI wavelengths were found by substituting into the fit equation for each wavelength on each phantom. Comparisons between the systems were made by averaging the pixels within a region of interest from each optical property map. The average optical properties for each of the three measurements is plotted in Fig. 4. The difference in between the openSFDI systems and the commercial system was , , and for the FLIR-, Andor-, and Thorlabs-based systems respectively, while the difference in was found to be , , and , respectively. As evidenced by the larger vertical error bars in the top row of Fig 4, the FLIR system had more variation than the Andor or Thorlabs systems. This variability is likely due to the fact that the FLIR sensor is not cooled. Bland–Altman analysis43 was performed to compare the FLIR openSFDI system to the commercial device. We found that the average difference between the two devices was in and in . Importantly, the accuracy was maintained across the entire range of and values tested. The limits of agreement in were and and the limits of agreement in were and , respectively. In Bland–Altman analysis, the limits of agreement indicate the region where difference between the two instruments will fall 95% of the time. That is, if 100 measurements were made on the same samples with openSFDI and the commercial instrument, the differences between about 95 of those measurements will fall within the limits of agreement. However, there is uncertainty associated with the limits of agreement themselves as indicated by the error bars in Fig. 5. The importance of this uncertainty is illustrated in Fig. 5 where there are 81 points per plot. In the absorption panel, eight points (9.8%) fall outside the estimated limits of agreement, while only three points (3.7%) fall outside the region bounded by the uncertainty of the limit of agreement estimates. Looking at the limits of agreement in isolation can give an inaccurate picture of the differences between the two instruments.44 5.2.PrecisionTo evaluate the precision of openSFDI, we conducted repeated measurements of the same optical phantom. Two different scenarios were explored. In the first scenario, SFDI measurements were repeated every 20 s over a 15-min period. This corresponds to a situation in which time resolution is important (e.g., measurements during a vascular occlusion). In the second scenario, SFDI measurements were repeated every minute for 1 h. This corresponds to a situation in which slower changes in optical properties are expected. Figure 6 demonstrates that there is no trend in either or that would indicate drift over time. Rather, the differences in and move randomly about the mean. Overall, the average standard deviation of the measurements across all wavelengths was in both the shorter drift measurement and the 1-h measurement. In , the average standard deviation was during the short duration measurement and over the 1-h measurement. 5.3.ExtensionsThe openSFDI system was designed to be easily modified and extended. Wherever possible, modular components were chosen to allow for modifications to the system to be made. In this section, we discuss several potential openSFDI modifications. 5.3.1.WavelengthsThe three wavelengths described in this work (660, 735, and 865 nm) were chosen for their availability and sensitivity to oxyhemoglobin and deoxyhemoglobin.45 However, different wavelengths can be used to target different chromophores. Altering the wavelengths of the openSFDI system can be accomplished by swapping one of the LED assemblies and (if necessary) changing the dichroic mirrors. 5.3.2.ProfilometryThe profile of the sample can be used to account for portions of the sample that are closer or farther away from the camera, providing more accurate optical properties for samples with irregularly geometry.46,47 Briefly, this is accomplished by collecting measurements of a phantom with known optical properties at different heights. From these measurements, a linear fit comparing pixel intensity to height can be derived for each pixel. Once the object’s profile is known, this linear fit can be used to add or subtract intensity based on the calibration measurement. Profilometry can be performed using a phase-stepping technique in which sinusoidal patterns are projected onto the sample. The unwrapped phase image maps directly to the object’s profile.46 Figure 7 shows an example of profilimitry conducted with the openSFDI system on a hemispheric phantom. 5.3.3.Optics selectionThe openSFDI system also provides flexibility in the optics that govern projection and detection of patterns. The suggested design results in a field of view that is . The field of view can be adjusted by changing the height of the system, increasing the diameter of the relay mirror, or changing the projection lens. The basic design calls for a standard CMOS camera that can be interfaced with a variety of lenses to adjust the field of view. In our lab, we have tested a variety of other cameras and imaging lenses. We have validated the performance with both CMOS and sCMOS cameras (Fig. 4) through a variety of lenses. We have also successfully modified the openSFDI system to use a microscope objective as the imaging lens. Changing the field of view will also impact the intensity of the light hitting the sample. To ensure good signal-to-noise ratio of the images, the LED power, exposure times, or aperture of the imaging lens may be necessary. 6.DiscussionThe openSFDI system described here and on the companion website provides a low-cost and step-by-step way for researchers to explore SFDI as a potential tool for new applications.28 However, as with any imaging modality, there are particular challenges that can limit the accuracy of SFDI imaging. While touched on in the body of the paper, there are a few points that deserve more careful attention. DMDs are inherently binary, with each mirror being switched fully on or fully off. Shades of gray are realized by pulse-width modulation (PWM) of the mirrors. The period of the PWM is known as the refresh rate of the DMD which is 16.7 ms for the LC4500. When imaging, care should be taken not to reduce the exposure time below the refresh rate as it will result in a distorted image. Also, we have observed in some SFDI systems that using an exposure time that is an integral multiple of the refresh rate can reduce the presence of artifacts in the demodulated images. However, no effort to match these values was made in this work. The openSFDI does not include source code for processing. This is by design so that users can gain experience in developing each step in the processing chain for their particular application. We note that special care is warranted in the matter of calibration. The accuracy and precision of the calibration phantom optical properties will affect subsequent downstream processing steps and overall SFDI results. Though it is beyond the scope of this paper, we encourage users to consider how potential errors in the calibration phantom optical properties propagate to the final images. 7.ConclusionIn this paper, we have outlined the design and performance of openSFDI, an open source hardware SFDI system. The companion website to this paper28 provides a complete materials list as well as step-by-step instructions for constructing and aligning the openSFDI platform. The website also provides software written in LabVIEW that enables rapid acquisition of SFDI datasets and detailed instructions for developing a data-processing pipeline. Here we described three different openSFDI systems built at two different locations, each of which had excellent accuracy and precision in optical property extractions and was able to measure oxyhemoglobin and deoxyhemoglobin concentrations in biological samples. We hope that the openSFDI platform will lower the barrier-to-entry for SFDI and help make it more accessible to researchers around the world. DisclosuresThe authors have no relevant financial interests in the article and no other potential conflicts of interest to disclose. AcknowledgmentsThe authors thank Evelyn Vu and Carlos Gomez for their help in collecting SFDI data. M.B.A., K.K., S.M.T., and D.R. gratefully acknowledge funding from the U.S. Department of Defense (DoD) (Award No. W81XWH-15-1-0070). R.B.S. gratefully acknowledges funding from the Knut and Alice Wallenberg Foundation. S.G. gratefully acknowledges funding from the European Research Council under the European Union’s Horizon 2020 Research and Innovation Program under Grant agreement No. 715737 and the French program “Investissement d’Avenir” run by the “Agence Nationale de la Recherche” under Grant Agreement No. ANR-11-INBS-006. K.K. gratefully acknowledges funding from the National Cancer Institute of the National Institutes of Health (NIH) (Award No. F99CA223014). The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH. K.T. and W.A. acknowledge funding from National Science Foundation # 1832886. ReferencesD. J. Cuccia et al.,
“Quantitation and mapping of tissue optical properties using modulated imaging,”
J. Biomed. Opt., 14
(2), 024012
(2009). https://doi.org/10.1117/1.3088140 JBOPFO 1083-3668 Google Scholar
A. Yafi et al.,
“Postoperative quantitative assessment of reconstructive tissue status in a cutaneous flap model using spatial frequency domain imaging,”
Plast. Reconstr. Surg., 127 117
–130
(2011). https://doi.org/10.1097/PRS.0b013e3181f959cc Google Scholar
A. Ponticorvo et al.,
“Evaluating visual perception for assessing reconstructed flap health,”
J. Surg. Res., 197 210
–217
(2015). https://doi.org/10.1016/j.jss.2015.03.099 JSGRA2 0022-4804 Google Scholar
C. R. Vargas et al.,
“Intraoperative hemifacial composite flap perfusion assessment using spatial frequency domain imaging: a pilot study in preparation for facial transplantation,”
Ann. Plastic Surg., 76 249
–255
(2016). https://doi.org/10.1097/SAP.0000000000000631 APCSD4 Google Scholar
A. J. Lin et al.,
“Spatial frequency domain imaging of intrinsic optical property contrast in a mouse model of Alzheimer’s disease,”
Ann. Biomed. Eng., 39 1349
–1357
(2011). https://doi.org/10.1007/s10439-011-0269-6 ABMECF 0090-6964 Google Scholar
A. M. Laughney et al.,
“Spectral discrimination of breast pathologies in situ using spatial frequency domain imaging,”
Breast Cancer Res., 15 R61
(2013). https://doi.org/10.1186/bcr3455 BCTRD6 Google Scholar
S. Nandy et al.,
“Characterizing optical properties and spatial heterogeneity of human ovarian tissue using spatial frequency domain imaging,”
J. Biomed. Opt., 21
(10), 101402
(2016). https://doi.org/10.1117/1.JBO.21.10.101402 JBOPFO 1083-3668 Google Scholar
C. K. Hayakawa et al.,
“Optical sampling depth in the spatial frequency domain,”
J. Biomed. Opt., 24
(7), 071603
(2018). https://doi.org/10.1117/1.JBO.24.7.071603 JBOPFO 1083-3668 Google Scholar
A. Mazhar et al.,
“Spatial frequency domain imaging of port wine stain biochemical composition in response to laser therapy: a pilot study,”
Lasers Surg. Med., 44
(8), 611
–621
(2012). https://doi.org/10.1002/lsm.22067 LSMEDI 0196-8092 Google Scholar
S. A. Sharif et al.,
“Noninvasive clinical assessment of port-wine stain birthmarks using current and future optical imaging technology: a review,”
Br. J. Dermatol., 167 1215
–1223
(2012). https://doi.org/10.1111/bjd.2012.167.issue-6 BJDEAZ 0007-0963 Google Scholar
D. M. Burmeister et al.,
“Utility of spatial frequency domain imaging (SFDI) and laser speckle imaging (LSI) to non-invasively diagnose burn depth in a porcine model,”
Burns, 41 1242
–1252
(2015). https://doi.org/10.1016/j.burns.2015.03.001 BURND8 0305-4179 Google Scholar
A. Ponticorvo et al.,
“Quantitative long-term measurements of burns in a rat model using spatial frequency domain imaging (SFDI) and laser speckle imaging (LSI),”
Lasers Surg. Med., 49
(3), 293
–304
(2017). https://doi.org/10.1002/lsm.22647 LSMEDI 0196-8092 Google Scholar
T. T. A. Nguyen et al.,
“Novel application of a spatial frequency domain imaging system to determine signature spectral differences between infected and noninfected burn wounds,”
J. Burn Care Res., 34
(1), 44
–50
(2013). https://doi.org/10.1097/BCR.0b013e318269be30 Google Scholar
T. E. Travis et al.,
“A multimodal assessment of melanin and melanocyte activity in abnormally pigmented hypertrophic scar,”
J. Burn Care Res., 36
(1), 77
–86
(2015). https://doi.org/10.1097/BCR.0000000000000154 Google Scholar
A. Yafi et al.,
“Quantitative skin assessment using spatial frequency domain imaging (SFDI) in patients with or at high risk for pressure ulcers,”
Lasers Surg. Med., 49 827
–834
(2017). https://doi.org/10.1002/lsm.v49.9 LSMEDI 0196-8092 Google Scholar
D. J. Rohrbach et al.,
“Characterization of nonmelanoma skin cancer for light therapy using spatial frequency domain imaging,”
Biomed. Opt. Express, 6
(5), 1761
–1766
(2015). https://doi.org/10.1364/BOE.6.001761 BOEICL 2156-7085 Google Scholar
C. Weinkauf et al.,
“Near-instant noninvasive optical imaging of tissue perfusion for vascular assessment,”
J. Vasc. Surg., 69 555
–562
(2019). https://doi.org/10.1016/j.jvs.2018.06.202 Google Scholar
S. Tabassum et al.,
“Feasibility of spatial frequency domain imaging (SFDI) for optically characterizing a preclinical oncology model,”
Biomed. Opt. Express, 7
(10), 4154
–4170
(2016). https://doi.org/10.1364/BOE.7.004154 BOEICL 2156-7085 Google Scholar
D. Hu et al.,
“Noncontact and wide-field characterization of the absorption and scattering properties of apple fruit using spatial-frequency domain imaging,”
Sci. Rep., 6 37920
(2016). https://doi.org/10.1038/srep37920 SRCEC3 2045-2322 Google Scholar
Y. Sun et al.,
“Detection of early decay in peaches by structured-illumination reflectance imaging,”
Postharvest Biol. Technol., 151 68
–78
(2019). https://doi.org/10.1016/j.postharvbio.2019.01.011 PBTEED Google Scholar
J. P. Angelo et al.,
“Review of structured light in diffuse optical imaging,”
J. Biomed. Opt., 24 071602
(2018). https://doi.org/10.1117/1.JBO.24.7.071602 JBOPFO 1083-3668 Google Scholar
S. Gioux, A. Mazhar and D. J. Cuccia,
“Spatial frequency domain imaging in 2019: principles, applications, and perspectives,”
J. Biomed. Opt., 24
(7), 071613
(2019). https://doi.org/10.1117/1.JBO.24.7.071613 JBOPFO 1083-3668 Google Scholar
M. B. Applegate,
“OpenSFDI Materials,”
(2019) www.openSFDI.org/materials December ). 2019). Google Scholar
S. Gioux et al.,
“First-in-human pilot study of a spatial frequency domain oxygenation imaging system,”
J. Biomed. Opt., 16
(8), 086015
(2011). https://doi.org/10.1117/1.3614566 JBOPFO 1083-3668 Google Scholar
R. B. Saager et al.,
“A light emitting diode (LED) based spatial frequency domain imaging system for optimization of photodynamic therapy of nonmelanoma skin cancer: quantitative reflectance imaging,”
Lasers Surg. Med., 45 207
–215
(2013). https://doi.org/10.1002/lsm.v45.4 LSMEDI 0196-8092 Google Scholar
J. P. Angelo, M. van de Giessen and S. Gioux,
“Real-time endoscopic optical properties imaging,”
Biomed. Opt. Express, 8 5113
–5126
(2017). https://doi.org/10.1364/BOE.8.005113 BOEICL 2156-7085 Google Scholar
M. B. Applegate and D. Roblyer,
“High-speed spatial frequency domain imaging with temporally modulated light,”
J. Biomed. Opt., 22 076019
(2017). https://doi.org/10.1117/1.JBO.22.7.076019 JBOPFO 1083-3668 Google Scholar
R. B. Saager et al.,
“Low-cost tissue simulating phantoms with adjustable wavelength-dependent scattering properties in the visible and infrared ranges,”
J. Biomed. Opt., 21
(6), 067001
(2016). https://doi.org/10.1117/1.JBO.21.6.067001 JBOPFO 1083-3668 Google Scholar
F. R. Ayers et al.,
“Wide-field spatial mapping of in vivo tattoo skin optical properties using modulated imaging,”
Lasers Surg. Med., 41 442
–453
(2009). https://doi.org/10.1002/lsm.v41:6 LSMEDI 0196-8092 Google Scholar
J.-P. Bouchard et al.,
“Reference optical phantoms for diffuse optical spectroscopy. Part 1—error analysis of a time resolved transmittance characterization method,”
Opt. Express, 18
(11), 11495
–11507
(2010). https://doi.org/10.1364/OE.18.011495 OPEXFF 1094-4087 Google Scholar
S. A. Prahl, M. J. C. van Gemert and A. J. Welch,
“Determining the optical properties of turbid media by using the adding-doubling method,”
Appl. Opt., 32
(4), 559
–568
(1993). https://doi.org/10.1364/AO.32.000559 APOPAI 0003-6935 Google Scholar
H. J. van Staveren et al.,
“Light scattering in intralipid-10% in the wavelength range of 400–1100 nm,”
Appl. Opt., 30
(31), 4507
–4514
(1991). https://doi.org/10.1364/AO.30.004507 APOPAI 0003-6935 Google Scholar
I. Driver et al.,
“The optical properties of aqueous suspensions of intralipid, a fat emulsion,”
Phys. Med. Biol., 34
(12), 1927
–1930
(1989). https://doi.org/10.1088/0031-9155/34/12/015 PHMBA7 0031-9155 Google Scholar
Institut national d’optique, “INO biomimic phantoms,”
(2019) https://www.ino.ca/en/biomimic/ December ). 2019). Google Scholar
V. Pera et al.,
“Optical property uncertainty estimates for spatial frequency domain imaging,”
Biomed. Opt. Express, 9 661
–678
(2018). https://doi.org/10.1364/BOE.9.000661 BOEICL 2156-7085 Google Scholar
K. P. Nadeau et al.,
“Quantitative assessment of renal arterial occlusion in a porcine model using spatial frequency domain imaging,”
Opt. Lett., 38 3566
–3569
(2013). https://doi.org/10.1364/OL.38.003566 OPLEDP 0146-9592 Google Scholar
K. P. Nadeau, A. J. Durkin and B. J. Tromberg,
“Advanced demodulation technique for the extraction of tissue optical properties and structural orientation contrast in the spatial frequency domain,”
J. Biomed. Opt., 19
(5), 056013
(2014). https://doi.org/10.1117/1.JBO.19.5.056013 JBOPFO 1083-3668 Google Scholar
R. B. Saager,
“OpenSFDI processing tutorial,”
(2019) www.openSFDI.org/processing December ). 2019). Google Scholar
W. G. Zijlstra, A. Buursma and W. P. Meeuwsen-van der Roest,
“Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin,”
Clin. Chem., 37
(9), 1633
–1638
(1991). Google Scholar
F. Ayers et al.,
“Fabrication and characterization of silicone-based tissue phantoms with tunable optical properties in the visible and near infrared domain,”
Proc. SPIE, 6870 687007
(2008). https://doi.org/10.1117/12.764969 PSISDG 0277-786X Google Scholar
A. R. Gardner and V. Venugopalan,
“Accurate and efficient Monte Carlo solutions to the radiative transport equation in the spatial frequency domain,”
Opt. Lett., 36
(12), 2269
–2271
(2011). https://doi.org/10.1364/OL.36.002269 OPLEDP 0146-9592 Google Scholar
J. M. Bland and D. G. Altman,
“Statistical methods for assessing agreement between two methods of clinical measurement,”
Lancet, 327
(8476), 307
–310
(1986). https://doi.org/10.1016/S0140-6736(86)90837-8 LANCAO 0140-6736 Google Scholar
C. Hamilton and J. Stamey,
“Using Bland-Altman to assess agreement between two medical devices—don’t forget the confidence intervals!,”
J. Clin. Monit. Comput., 21
(6), 331
–333
(2007). https://doi.org/10.1007/s10877-007-9092-x Google Scholar
A. Mazhar et al.,
“Wavelength optimization for rapid chromophore mapping using spatial frequency domain imaging,”
J. Biomed. Opt., 15
(6), 061716
(2010). https://doi.org/10.1117/1.3523373 JBOPFO 1083-3668 Google Scholar
S. Gioux et al.,
“Three-dimensional surface profile intensity correction for spatially-modulated imaging,”
J. Biomed. Opt., 14
(3), 034045
(2009). https://doi.org/10.1117/1.3156840 JBOPFO 1083-3668 Google Scholar
Y. Zhao et al.,
“Angle correction for small animal tumor imaging with spatial frequency domain imaging (SFDI),”
Biomed. Opt. Express, 7
(6), 2373
–2384
(2016). https://doi.org/10.1364/BOE.7.002373 BOEICL 2156-7085 Google Scholar
BiographyMatthew B. Applegate is a postdoctoral fellow in Darren Roblyer’s lab at Boston University where his work focuses on developing new diffuse optical imaging technologies. He received his BS from Cornell University, and his PhD in biomedical engineering from Tufts University, where he studied under professor Fiorenzo Omenetto. His PhD research was centered around developing methods for the photomodification of biomaterials, especially silk fibroin, for biomedical applications. Kavon Karrobi is a PhD candidate in the Department of Biomedical Engineering at Boston University. He received his BS degree in biomedical engineering from Boston University in 2014. Currently, he is a F99 fellow of the National Cancer Institute of the National Institutes of Health. Karissa Tilbury is an assistant professor of biomedical engineering at the University of Maine. She received her BS (2009), MS (2011), and PhD (2015) in biomedical engineering from the University of Wisconsin. She did her postdoctoral work at the Biophotonics Center at Vanderbilt University. Rolf B. Saager is an assistant professor in the Department of Biomedical Engineering, Linköping University, Sweden, and a research fellow at the Wallenberg Centre for Molecular Medicine. His research focus is in biomedical imaging and spectroscopy. Prior to this position, he was a project scientist at Beckman Laser Institute and Medical Clinic, University of California, Irvine, and received his PhD in optics from the University of Rochester. Darren Roblyer is an associate professor in the Department of Biomedical Engineering at Boston University. His research is based on the development of diffuse optical spectroscopy and optical microscopy techniques. He received his BS degree in biomedical engineering from Johns Hopkins University in 2004, and received his PhD in 2009 from the Bioengineering Department at Rice University where he studied under Professor Rebecca Richards-Kortum. He was a Department of Defense postdoctoral fellow at the Beckman Laser Institute at the University of California, Irvine, studying under Professor Bruce Tromberg, prior to joining the faculty at Boston University in 2012. |