1.IntroductionBiomedical sensing and imaging play a crucial role in modern clinical analysis, providing visualizations of body tissues, facilitating non-invasive or minimally invasive diagnostic information, and offering real-time surgical guidance.1,2 It helps to reduce the time and cost of diagnosis and treatment and improve patient satisfaction. However, the lack of standards for evaluating devices and comparing systems across the field hampers both technological advancement and the clinical translation of optical sensing and imaging systems.3,4 Hemoglobin concentration in human blood is an important parameter to monitor the physiological condition. It gives information on the capability of oxygen transportation in blood, provides preoperative and postoperative care, helps to diagnose diseases such as anemia, evaluates treatment response, and assesses the performance of athletes.5–10 With the emergence of portable and wearable health monitoring devices, there is an increasing demand for optical standards capable of dynamically changing their optical properties to replicate changes in hemoglobin oxygen saturation during the respiratory and heart-beat cycle. Biophotonics phantoms are artificial structures designed to simulate the optical properties of biological media. They are essential tools for simulating light distribution with the geometry of physical tissues, validating and characterizing the performance of optical devices, and recording reference measurements.11 Many different phantoms with high stability, controllability, and repeatability have been developed for device calibration, optimization, and quality control.12–16 Researchers have developed liquid phantoms using whole blood or extracted erythrocytes to mimic changes in oxygen saturation. However, the use of blood components could raise ethical and safety concerns and limit the stability and shelf life of the phantoms. In addition, biological variability in blood components reduces the reproducibility of the phantom. To maintain the appropriate conditions, temperature and pH must be monitored throughout the entire measurement process. The use of temperature and pH monitors, peristaltic pumps, and membrane oxygenators increases the cost and complexity of the experimental setup. Oxygen saturation levels are modulated by adding oxygenating compounds such as oxygen and deoxygenating compounds such as yeast or nitrogen bubbling or a mixture of d-glucose, glucose oxidase, and catalase.17–23 A peristaltic pump is used for blood circulation, and a membrane oxygenator is applied to control blood oxygenation. Yeast takes 20 to 30 min to gradually deoxygenate the phantom,21 and nitrogen bubbling takes 30 min to effectively deoxygenate the phantom,22 which limits the potential for real-time control. Several solid dynamic phantoms have been developed to control changes in the absorption coefficient with high stability and reproducibility. Koh et al.24 designed a dynamic phantom with a liquid-crystal display (LCD) sandwiched between the solid tissue-mimicking phantom. The absorption at each pixel is electronically controlled to be on or off state. Funane et al.25 developed a phantom with two stage-driven absorber layers to realize gradual absorption changes by tuning the position of the absorber. Hebden et al.26 proposed a phantom with embedded targets comprising a black thermochromic pigment. The pigment changes from black to white when heated to 47°C. However, the performance of the LCD-based dynamic phantom was largely influenced by the scattering properties of the top layer phantoms because the attenuation of the LCD is angular dependent. The phantom with moving stages only creates macroscopic changes in tissue, and it lacks the potential to simulate complex structures such as vascular systems. For the thermochromic dynamic phantom, the smallest spacing between targets is 11.2 mm to avoid crosstalk from the neighboring target, thereby limiting the spatial resolution. In addition, it takes to change the color from room temperature, which is not appropriate for fast control. In this work, we implemented a novel light guide-based dynamic phantom to overcome the challenges mentioned above.27 This phantom utilized a polymer optical fiber to isolate the LCD from the top-layer tissue-mimicking phantom and avoid angular-dependent performance. The absorption was tuned gradually by applying different voltages to the LCD. The phantom was characterized at four wavelengths: 455, 530, 660, and 940 nm, covering both visible and near-infrared ranges. As case studies, photoplethysmography (PPG) signals of different heart rates and oxygen saturation levels were generated by the dynamic phantom and tested by a photodiode. This dynamic phantom can be a helpful tool for calibrating and validating pulse oximeters, imaging systems, and spectroscopy devices. 2.Method and Material2.1.Phantom DesignFigure 1(a) shows the photo of the dynamic phantom. Figure 1(b) illustrates the schematic layout of the dynamic phantom. The light passing through a static tissue-mimicking silicone phantom was collected by a 4-cm long 3-mm diameter polymer optical fiber (OMPF3000, The Optoelectronic Manufacturing Corporation Ltd., Redruth, United Kingdom) and collimated by a condenser lens (ACL2520U, Thorlabs, Newton, New Jersey, United States) to avoid the angular dependent performance of the LCD. The parallel beam was modulated by the LCD (FOS-NIR (1100), LC-Tec Displays AB, Borlänge, Sweden) and then reflected through the phantom by a silver mirror (PF10-03-P01, Thorlabs). The LCD was controlled by a National Instruments data acquisition (NI DAQ) (USB-6212, National Instrument, Austin, Texas, United States). The thickness and optical properties of the static superficial silicone phantom can vary depending on the application. In our demonstration, a 9-cm diameter, 0.5-mm-thick round shape silicone phantom was used based on the recipe described in Ref. 15. Briefly, the silicone phantom consisted of SiliGlass (MB Fibreglass, Newtownabbey, United Kingdom) as bulk material, polycraft black silicone pigment (MB Fibreglass, UK) as absorber, and silica microspheres (440345, Sigma-Aldrich, St. Louis, Missouri, United States) as a scatterer. The optical properties of this silicone phantom were characterized by a time-domain diffuse optical spectrometer.28,29 The target absorption coefficient () was , and the reduced scattering coefficient () was . 2.2.Phantom Characterization SystemThe full dynamic phantom was further characterized by an in-house system described in Fig. 1(c). The illumination part of the system consisted of four light-emitting diodes (LEDs) at 455, 530, 660, and 940 nm (LED1: M530L4, LED2: M455L4, LED3: M660L4, and LED4: M940L4, Thorlabs). Dichroic mirrors [D1: 490-nm longpass dichroic mirror (DMLP490R, Thorlabs), D2: 605-nm shortpass dichroic mirror (DMSP605R, Thorlabs), D3: 805-nm shortpass dichroic mirror (DMSP805R, Thorlabs)] were employed to combine these LEDs into one single beam. The illumination wavelengths were selected by switching on or off any LED or any combination of LEDs. The reflected light from the phantom above the light guide was collected by a two-lens system (L1: AC254-050-AB-ML, L2: LB1676-ML, Thorlabs) and measured by a photodiode (S3590-08, Hamamatsu Photonics, Shizuoka, Japan) with an amplifier (DLPCA-200, FEMTO Messtechnik GmbH, Berlin, Germany) and fed into the data acquisition (DAQ) (USB-6212, National Instrument). 2.3.Dynamic Signal SynthesisThe process flow of phantom calibration is schematically described in Fig. 2(a). The intensity of the diffusely reflected light from the dynamic phantom was characterized by applying a voltage over the LCD from 0.5 to 3 V with a 0.05 V step size. The range in which this intensity was negatively correlated to the applied voltage was determined as the functional range. The response of this functional range was normalized to the maximum value to create a lookup table. The flowchart of the synthesis of the target signal is illustrated in Fig. 2(b). The target signal was also normalized to its maximum value to allow it to be mapped to the lookup table to get the modulated signal for the dynamic phantom. 2.4.Photoplethysmography SimulationTwo application case studies were implemented to demonstrate the dynamic phantom’s ability to reproduce different PPG signals at multiple wavelengths. One case study focused on generating PPG signals at various heart rates, whereas the other showcased the generation of PPG signals at different oxygen saturation levels. For the heart rate simulation case study, the target synthetic PPG signals with heart rates ranging from 80 to 120 beats per minute (bpm) in steps of 10 bpm were generated using the Python toolbox NeuroKit2.30 Each PPG signal contains 20,000 data points (20 s) sampled at 1000 Hz. The phantom modulation voltage signals were generated by mapping synthetic PPG signals to the lookup table, as described in Sec. 2.3. The dynamic phantom was modulated and measured by the phantom characterization system as described in Sec. 2.2 at 50-Hz sampling rate. The measured signals were then processed by NeuroKit2 to recover the heart rate.30 For the oxygen saturation case study, the wavelengths 660 and 940 nm were selected for the characterization to enable an evaluation of different oxygen saturation levels as these wavelengths are commonly used in commercial medical pulse oximeters. A single measured PPG signal recording of 6 s from a healthy volunteer31 was used. It was normalized to the maximum value. Targeted signal traces for two selected wavelengths and five tested blood oxygenation levels were then generated from this normalized single trace by adjusting the modulation depth according to the blood attenuation. These traces were then mapped to the lookup table to generate the modulated voltage signal for the dynamic phantom input. The phantom was measured by using the phantom characterization system as described in Sec. 2.2 at a 50-Hz sampling rate. The oxygen saturation can be obtained from measured PPG signals at these wavelengths through a double ratio of the pulsatile and non-pulsatile components of the measured red and near-infrared light32,33 where is the maximum amplitude of the pulse at 660 nm, is the minimum amplitude of the pulse at 660 nm, is the maximum amplitude of the pulse at 940 nm, and is the minimum amplitude of the pulse at 940 nm.In the region of 86% to 100%, oxygen saturation () shows a linear correlation with the rate34 The rate can thus simply be calculated by The linear range was simulated in this case study. The rate was calculated at every pulse, and then, the values were averaged over a 6-s sample. Target signals of different wavelengths and oxygen saturation levels were measured one by one. The relative change in the absorption coefficient of the arterial component of blood at 660 nm is 130% when changing from 100% to 86% oxygen saturation level, whereas at 940 nm, the relative absorption coefficient variation is only 6%.35 As the extinction coefficient of blood does not change significantly from 86% to 100% at 940 nm, we kept the amplitude constant at 940 nm and modulated the amplitude at 660 nm to simulate level at 86%, 90%, 95%, and 100%. 3.Results3.1.Dynamic Phantom CharacterizationThe results of the full dynamic phantom characterization are presented in Fig. 3(a). Here, the diffusely reflected light intensity from the phantom was measured at four wavelengths as a function of the voltage applied over the LCD. Within a certain voltage range (1.2 to 2.4 V), this intensity was negatively correlated to the applied voltage. This voltage range was identified as the functional range, as illustrated in Fig. 3(b). The lookup tables formed after normalizing the curve within the functional range for each wavelength are shown in Fig. 3(c). The maximal modulation depths are 7% at 940 nm, 13% at 660 nm, 8% at 530 nm, and 2% at 455 nm. The measured optical properties of the static top-layer silicone phantom were found to be , at 660 nm, and , at 940 nm, close to the targeted values. 3.2.Photoplethysmography Signals SynthesisNext, we demonstrate the capability of the dynamic phantom to replicate a targeted PPG trace using 660 nm as an example. Figure 4 shows the steps taken to generate a PPG trace at 660 nm, using the above-mentioned experimentally measured trace from Ref. 31 as a target [Fig. 4(a)]. The normalized diffuse reflectance response within the functional range of the dynamic phantom at 660 nm is shown in Fig. 4(b). The resulting modulation signal required for the LCD to generate the targeted diffuse reflectance from the dynamic phantom is presented in Fig. 4(c). This trace is generated by mapping the normalized original signal to the generated lookup table. Figure 4(d) shows the reproduced PPG signals measured by the photodiode at 660 nm, whereas the percentage error and average percentage error between the target signal [Fig. 4(a)] and the measured signal after max normalization of the target signal are shown in Fig. 4(e). The average percentage errors are 0.03% at 940 nm, 0.03% at 660 nm, 0.05% at 530 nm, and 0.3% at 455 nm. Measured diffuse reflectance traces at four wavelengths are given in Fig. 4(f). The results demonstrate that the dynamic phantom can successfully simulate PPG signals at multiple wavelengths. To demonstrate the phantom’s ability to generate dynamic biological signals, two case studies were conducted to generate PPG signals of different heart rates and oxygen saturation levels at multiple wavelengths. 3.3.Case Study 1: PPG Signals of Different Heart RatesFigure 5(a) shows 20-s PPG-like signal traces measured from the dynamic phantom for heart rates of 80, 90, 100, 110, and 120 bpm at 940 nm. Each trace shows clear periodicity corresponding to the designated heart rates, indicating the ability of the dynamic phantom to realistic cardiac rhythms. Figures 5(b)–5(e) compare the expected heart rates to the measured heart rates at 940, 660, 530, and 455 nm. The values of the linear regression at these four wavelengths are 0.999. The good agreement between the expected and measured values validates the accuracy of the dynamic phantom in replicating various heart rates, demonstrating its potential for calibrating and validating heart rate monitors. 3.4.Case Study 2: PPG Signal of Different Oxygen Saturation LevelsFigure 6(a) shows the target signals at different oxygen saturation levels. Figure 6(b) illustrates the measurement results at different oxygen saturation levels. The measured signals demonstrate clear amplitude variations that correlate with the specified oxygen saturation levels. Figure 6(c) presents the measured -value as a function of the expected levels, indicating a linear decrease in the -value with increasing . This result aligns with the conclusion of Ref. 35, which states that the -value should be 1.12, 0.96, 0.76, and 0.55 at 86%, 90%, 95%, and 100% oxygen saturation levels, respectively. Figure 6(d) shows the measured against the expected , confirming the accuracy of the dynamic phantom in simulating realistic levels. These results validate the capability of the dynamic phantom to accurately mimic hemodynamics signals and its potential application in calibrating and validating pulse oximeters. 4.DiscussionTwo case studies mentioned above demonstrate the ability of the dynamic phantom to generate PPG signals at different wavelengths and different oxygen saturation levels. For the characterization of the phantom, the functional range and contrast vary for different wavelengths due to the wavelength-dependent performance of the LCD. The operational wavelength range of the dynamic phantom is thereby determined by the specifications of the LCD. In this setup, the LCD provides broadband visual to near-infrared (400 to 2000 nm) operation with high open-state transmittance optimized for wavelengths around 1100 nm. Consequently, the LCD response shows a larger contrast at 940, 660, and 530 nm at 455 nm. It is also worth noting that only a small fraction of the dynamic range of the functional voltage range has been used to generate the modulation depth in the targeted PPG trace. The maximal modulation depth at 455 nm is 2%, whereas the perfusion index of the PPG signal from a health volunteer is around 1%.31 The noise levels are similar at all wavelengths when simulating a 1% perfusion index. The modulation frequency of the phantom is limited by the operation rates of the LCD (1000 Hz) and NI DAQ (400 kHz). In this study, a 50-Hz modulation frequency was used to match the sampling rate in Ref. 31. However, with an upper modulation frequency limit of 1000 Hz, the device can simulate any feasible human heart rate. This allows accurate reproduction of PPG pulse shape features, including the systolic peak, diastolic peak, and dicrotic notch. These capabilities highlight the phantom’s potential to simulate diverse PPG signal shapes for pulse shape analysis in cuffless blood pressure monitoring applications.36 In addition to the PPG signal, other dynamic variations in biological tissue, such as hemodynamics, can be simulated by applying the corresponding modulation signals. Compared with the phantom described in Ref. 24, which only characterizes spatial differentiation in attenuation for optical topography, our phantom demonstrates the ability to gradually modulate attenuation with a step size voltage modulation and simulate realistic biological signals, making it a more flexible tool for various biomedical applications. When simulating the PPG signals, various noise signals can be added to simulate motion artifacts introduced by physical movement during the measurement or low-intensity signals due to low blood perfusion. If the phantom is tested using a commercial pulse oximeter, the transmission configuration should be used instead of the reflection configuration. Also, the illumination and phantom modulation must be synchronized as the typical pulse oximeter quickly switches between two wavelengths.32 This can be achieved by having a sync input from the pulse oximeter to the phantom, making it possible to dynamically change the phantom modulation to the corresponding wavelength of the pulse oximeter. The other approach is to have two individual light guides and LCDs, one for 660-nm modulation and the other for 940-nm modulation, separately. In addition to generating dynamic signals, this setup could be used to characterize the spatial resolution of imaging or spectroscopy systems by varying the diameter of the light guide. Multiple light guides can be integrated to create specific patterns to mimic complex biological structures, such as vascular systems. There is no minimum space requirement between neighboring light guides. Compared with the thermochromic dynamic phantom described in Ref. 26, which has a target size of 4.5 mm and requires 11.2-mm spacing between targets, and the phantom presented in Ref. 25, which only provides macroscopic absorption changes, our phantom shows better potential for simulating small and structured biological features with various target sizes. In future research, we intend to incorporate an additional thin-layer phantom with varying melanin concentrations to create a multilayer phantom. This multilayer phantom would mimic different skin colors, facilitating the characterization of skin color bias in pulse oximeters.37 The phantom’s low cost makes it widely applicable for both industrial and academic research. The estimated bill of material cost of our demonstration setup is euros, although this may vary depending on the specifications of the LCD and NI DAQ. Currently, the dynamic phantom cannot simulate vascular compliance during the cardiac cycle due to its fixed geometric structure. To address this limitation, future enhancements could incorporate fiber bundles to mimic the vascular structure. The geometry of the dynamic area of the phantom can be controlled by activating or deactivating specific fibers inside the fiber bundle. 5.ConclusionIn this study, we have reported the design, characterization, and potential applications of a dynamic phantom capable of generating dynamic biological signals. The absorption properties of the phantom can be controlled in real time by modulating the LCD. The phantom was characterized at four wavelengths: 940, 660, 530, and 455 nm. This dynamic phantom can simulate various biological signals by applying corresponding modulation signals. Clinically relevant case studies were conducted to prove the ability to generate PPG signals of different heart rates and oxygen saturation levels. This dynamic phantom shows the potential of calibrating and validating pulse oximeter, imaging, and spectroscopy systems. DisclosuresSanathana Konugolu Venkata Sekar and Stefan Andersson-Engels are shareholders of BioPixS Ltd. with an interest in tissue optical phantoms. The other authors declare no conflicts of interest. Code and Data AvailabilityThe code used in this study is available on GitHub: https://github.com/Hui0930/Dynamic-phantom.git. AcknowledgmentsThe research work was supported by the Science Foundation Ireland IPIC Biomedical theme (Grant Nos. SFI-12/RC/2276_P2 and SFI/22/RP-2TF/10293), the European Research Council (ERC) (Grant No. 101164780), the European projects VASCOVID (Grant No. 101016087), and TinyBrains (Grant No. 101017113). We would like to acknowledge Konstantin Grygoryev for his assistance in proofreading this paper. ReferencesQ. Li et al.,
“Review of spectral imaging technology in biomedical engineering: achievements and challenges,”
J. Biomed. Opt., 18
(10), 100901 https://doi.org/10.1117/1.JBO.18.10.100901 JBOPFO 1083-3668
(2013).
Google Scholar
Y. Hoshi and Y. Yamada,
“Overview of diffuse optical tomography and its clinical applications,”
J. Biomed. Opt., 21
(9), 091312 https://doi.org/10.1117/1.JBO.21.9.091312 JBOPFO 1083-3668
(2016).
Google Scholar
L. Hacker et al.,
“Criteria for the design of tissue-mimicking phantoms for the standardization of biophotonic instrumentation,”
Nat. Biomed. Eng., 6
(5), 541
–558 https://doi.org/10.1038/s41551-022-00890-6
(2022).
Google Scholar
W. S. Tummers et al.,
“Regulatory aspects of optical methods and exogenous targets for cancer detection,”
Cancer Res., 77
(9), 2197
–2206 https://doi.org/10.1158/0008-5472.CAN-16-3217 CNREA8 0008-5472
(2017).
Google Scholar
C. D. Karakochuk et al.,
“Measurement and interpretation of hemoglobin concentration in clinical and field settings: a narrative review,”
Ann. N. Y. Acad. Sci., 1450
(1), 126
–146 https://doi.org/10.1111/nyas.14003 ANYAA9 0077-8923
(2019).
Google Scholar
G. Lee, S. Choi and K. Kim,
“Association of hemoglobin concentration and its change with cardiovascular and all-cause mortality,”
J. Am. Heart Assoc., 7
(3), e007723 https://doi.org/10.1161/JAHA.117.007723
(2018).
Google Scholar
S.-K. Park et al.,
“Noninvasive hemoglobin monitoring for maintaining hemoglobin concentration within the target range during major noncardiac surgery: a randomized controlled trial,”
J. Clin. Anesth., 93 111326 https://doi.org/10.1016/j.jclinane.2023.111326 JCLBE7
(2024).
Google Scholar
J. Kraitl et al.,
“Non-invasive measurement of blood components: sensors for an in-vivo haemoglobin measurement,”
Advancement in Sensing Technology: New Developments and Practical Applications, 237
–262 Springer, Berlin
(2013). Google Scholar
J. A. Calbet et al.,
“Importance of hemoglobin concentration to exercise: acute manipulations,”
Respir. Physiol. Neurobiol., 151
(2–3), 132
–140 https://doi.org/10.1016/j.resp.2006.01.014 RPNEAV 1569-9048
(2006).
Google Scholar
L. M. Neufeld et al.,
“Hemoglobin concentration and anemia diagnosis in venous and capillary blood: biological basis and policy implications,”
Ann. N. Y. Acad. Sci., 1450
(1), 172
–189 https://doi.org/10.1111/nyas.14139 ANYAA9 0077-8923
(2019).
Google Scholar
S. Prahl,
“Project: optical phantoms,”
https://omlc.org/~prahl/projects/phantoms.html
().
Google Scholar
M. Z. Vardaki and N. Kourkoumelis,
“Tissue phantoms for biomedical applications in Raman spectroscopy: a review,”
Biomed. Eng. Comput. Biol., 11 1179597220948100 https://doi.org/10.1177/1179597220948100
(2020).
Google Scholar
L. Ntombela, B. Adeleye and N. Chetty,
“Low-cost fabrication of optical tissue phantoms for use in biomedical imaging,”
Heliyon, 6
(3), e03602 https://doi.org/10.1016/j.heliyon.2020.e03602
(2020).
Google Scholar
B. W. Pogue and M. S. Patterson,
“Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry,”
J. Biomed. Opt., 11
(4), 041102 https://doi.org/10.1117/1.2335429 JBOPFO 1083-3668
(2006).
Google Scholar
S. K. V. Sekar et al.,
“Solid phantom recipe for diffuse optics in biophotonics applications: a step towards anatomically correct 3D tissue phantoms,”
Biomed. Opt. Express, 10
(4), 2090
–2100 https://doi.org/10.1364/BOE.10.002090 BOEICL 2156-7085
(2019).
Google Scholar
A. B. Walter and E. D. Jansen,
“Development of a platform for broadband, spectra-fitted, tissue optical phantoms,”
J. Biomed. Opt., 28
(2), 025001 https://doi.org/10.1117/1.JBO.28.2.025001 JBOPFO 1083-3668
(2023).
Google Scholar
S. Kleiser et al.,
“Comparison of tissue oximeters on a liquid phantom with adjustable optical properties,”
Biomed.Opt. Express, 7
(8), 2973
–2992 https://doi.org/10.1364/BOE.7.002973 JBOPFO 1083-3668
(2016).
Google Scholar
H. Isler et al.,
“Liquid blood phantoms to validate NIRS oximeters: yeast versus nitrogen for deoxygenation,”
Oxygen Transport to Tissue XL, 381
–385 Springer, Cham
(2018). Google Scholar
X. Lv et al.,
“Design of a portable phantom device to simulate tissue oxygenation and blood perfusion,”
Appl. Opt., 57
(14), 3938
–3946 https://doi.org/10.1364/AO.57.003938 APOPAI 0003-6935
(2018).
Google Scholar
C. D. Kurth et al.,
“A dynamic phantom brain model for near-infrared spectroscopy,”
Phys. Med. Biol., 40
(12), 2079 https://doi.org/10.1088/0031-9155/40/12/006 PHMBA7 0031-9155
(1995).
Google Scholar
E. L. Hull, M. G. Nichols and T. H. Foster,
“Quantitative broadband near-infrared spectroscopy of tissue-simulating phantoms containing erythrocytes,”
Phys. Med. Biol., 43
(11), 3381 https://doi.org/10.1088/0031-9155/43/11/014 PHMBA7 0031-9155
(1998).
Google Scholar
T. O. McBride et al.,
“Spectroscopic diffuse optical tomography for the quantitative assessment of hemoglobin concentration and oxygen saturation in breast tissue,”
Appl. Opt., 38
(25), 5480 https://doi.org/10.1364/AO.38.005480 APOPAI 0003-6935
(1999).
Google Scholar
R. Zhang et al.,
“Čerenkov radiation emission and excited luminescence (CREL) sensitivity during external beam radiation therapy: Monte Carlo and tissue oxygenation phantom studies,”
Biomed. Opt. Express, 3
(10), 2381 https://doi.org/10.1364/BOE.3.002381 BOEICL 2156-7085
(2012).
Google Scholar
P. H. Koh, C. E. Elwell, D. T. Delpy,
“Development of a dynamic test phantom for optical topography,”
Oxygen Transport to Tissue XXX, 141
–146 Springer, Boston, MA
(2009). Google Scholar
T. Funane et al.,
“Dynamic phantom with two stage-driven absorbers for mimicking hemoglobin changes in superficial and deep tissues,”
J. Biomed. Opt., 17
(4), 047001 https://doi.org/10.1117/1.JBO.17.4.047001 JBOPFO 1083-3668
(2012).
Google Scholar
J. C. Hebden et al.,
“An electrically-activated dynamic tissue-equivalent phantom for assessment of diffuse optical imaging systems,”
Phys. Med. Biol., 53
(2), 329
–337 https://doi.org/10.1088/0031-9155/53/2/002 PHMBA7 0031-9155
(2007).
Google Scholar
H. Ma et al.,
“Light-guided dynamic phantom to mimic microvascular for biomedical applications,”
in Opt. Biophotonics Congr.: Biomed. Opt. 2024 (Transl., Microsc., OCT, OTS, BRAIN),
(2024). https://doi.org/10.1364/ots.2024.ow3d.2 Google Scholar
S. Konugolu Venkata Sekar et al.,
“Diffuse optical characterization of collagen absorption from 500 to 1700 nm,”
J. Biomed. Opt., 22
(1), 015006 https://doi.org/10.1117/1.JBO.22.1.015006 JBOPFO 1083-3668
(2017).
Google Scholar
S. Konugolu Venkata Sekar et al.,
“Broadband time domain diffuse optical reflectance spectroscopy: a review of systems, methods, and applications,”
Appl. Sci., 9
(24), 5465 https://doi.org/10.3390/app9245465
(2019).
Google Scholar
D. Makowski et al.,
“NeuroKit2: a Python toolbox for neurophysiological signal processing,”
Behav. Res. Methods, 53
(4), 1689
–1696 https://doi.org/10.3758/s13428-020-01516-y
(2021).
Google Scholar
A. I. Siam et al.,
“Real-world PPG dataset,”
(2022). Google Scholar
E. D. Chan, M. M. Chan, M. M. Chan,
“Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations,”
Respir. Med., 107
(6), 789
–799 https://doi.org/10.1016/j.rmed.2013.02.004 RMEDEY 0954-6111
(2013).
Google Scholar
S. Kästle et al.,
“A new family of sensors for pulse oximetry,”
Hewlett Packard J., 48 39
–47
(1997).
Google Scholar
N. Stubán and N. Masatsugu,
“Non-invasive calibration method for pulse oximeters,”
Period. Polytech. Electr. Eng., 52
(1–2), 91
–94 https://doi.org/10.3311/pp.ee.2008-1-2.11 PPYTA7
(2008).
Google Scholar
S. Prahl,
“Summary of hemoglobin spectra,”
https://omlc.org/spectra/hemoglobin/summary.html
().
Google Scholar
J.-R. Hu et al.,
“Validating cuffless continuous blood pressure monitoring devices,”
Cardiovasc. Digit. Health J., 4
(1), 9
–20 https://doi.org/10.1016/j.cvdhj.2023.01.001
(2023).
Google Scholar
R. Al-Halawani et al.,
“A review of the effect of skin pigmentation on pulse oximeter accuracy,”
Physiol. Meas., 44
(5), 05TR01 https://doi.org/10.1088/1361-6579/acd51a PMEAE3 0967-3334
(2023).
Google Scholar
BiographyHui Ma is a PhD student in the Biophotonics group at Tyndall National Institute, Cork, Ireland. Their research focuses on Raman Spectroscopy, beam shaping, phantoms, and standardization, contributing to advancements in optical diagnostic techniques. She received her bachelor’s degree in optoelectrical information and engineering from Huazhong University of Science and Technology and her master’s degree in photonics integration from Eindhoven University of Technology. Dario Angelone is a PhD student in the Biophotonics group at Tyndall National Institute, Cork, Ireland. His work focuses on Fluorescence Lifetime Imaging (FLIM) for biomedical applications and is part of the PHAST project funded by the EU under Horizon 2020. He received his bachelor’s degree in physics from the University of Aberdeen before being awarded an Erasmus Mundus scholarship for the European joint master’s degree in Smart Systems Integration at Heriot-Watt University Edinburgh. Claudia Nunzia Guadagno is CTO at BioPixS Ltd. She received her master’s degree in physics Engineering in 2016. Her thesis topic was in time-domain diffuse optics for in vivo and phantom application. Her area interests are time domain diffuse optics, phantoms and standardization in biophotonics. She formerly worked for different companies and also received a BSc degree in Communication in 2007. Stefan Andersson-Engels has led the biophotonics group at the Irish Photonic Integration Centre (IPIC) since 2016. He received his PhD in physics from Lund University in 1990, following a degree in Engineering Physics, and was a postdoc at McMaster University. Appointed professor in 1999, he has published extensively in biomedical optics, focusing on optical spectroscopy and photodynamic therapy, as well as light propagation in turbid media. He is a senior advisory editor for the Journal of Biomedical Optics. Sanathana Konugolu Venkata Sekar is biomedical theme coordinator at IPIC and an associate professor at University College Cork. He is head of the Integrated FAST Biophotonics Group at Tyndall and co-founded BioPixS Ltd. A recipient of the ERC award, he oversees multidisciplinary research and supervises a team of PhD students and postdocs. His interests include phantoms, standardization, tissue oximeters, time-domain spectroscopy and multimodal microscopy, with a focus on translating photonics into miniaturized healthcare solutions. |
Liquid crystal displays
Modulation
Oxygen
Heart
Oximeters
Signal generators
Biomedical optics