1.BackgroundPulse oximetry is a well-established technique to estimate arterial oxygen saturation () from measurements of oxygen saturation () obtained from pulsatile optical signals at red (typically ) and infrared (IR) (typically ) wavelengths. The “p” in indicates that its measurement is based on the arterial “pulse,” and it also points to the fact that it pertains to “peripheral” blood. Many helpful reviews have been written about pulse oximetry, including one that recounts its history,1 one written in a way to convey the engineering principles to medical practitioners,2 one that discusses the current technology and its limitations,3 and one that provides a summary of the basic principles and possible sources of error.4 Pulse oximeters operate in a diffuse optical regime. Another diffuse optical technique is near-infrared spectroscopy (NIRS), which is aimed more at bulk tissue properties and hemodynamics, as opposed to pulse oximetry, which specifically targets the arterial vascular compartment. The ability to perform absolute measurements with NIRS depends on the temporal domain employed: continuous-wave (CW),5 frequency-domain (FD),6 or time-domain (TD)7 (listed in order of increasing information provided about the investigated tissue). FD and TD are capable of absolute quantification of bulk tissue properties, whereas CW typically is only able to recover relative hemodynamics or oxygenation changes. CW NIRS methods are conceptually similar to those used in pulse oximetry. NIRS has been broadly applied to various human tissues, including the brain,8 skeletal muscle,9 and breast.10 Each application of NIRS centers around either the absolute optical properties of tissues or their dynamic changes, which are then linked to associated anatomical or physiological information. As pulse oximetry is entirely focused on the temporal dynamics of optical signals, we will only consider this type of application. In NIRS, temporal dynamics are analyzed with the modified the Beer–Lambert law (mBLL), which translates an intensity change () into an absorption coefficient change () in tissue.11 At red and IR wavelengths, this typically results from a change in oxy-hemoglobin concentration () and a change in deoxy-hemoglobin concentration () in tissue, so that mBLL is used to retrieve hemodynamic information. Hemodynamic information is also what is sought in pulse oximetry, where and result from a pulsatile arterial blood-volume change (). This perspective serves to connect common data analysis methods in pulse oximetry to general NIRS concepts, which are formalized around the mBLL. In particular, we will discuss about possible interpretations of mBLL applied to pulse oximetry. The philosophy of the methods used for pulse oximetry and NIRS is not the same. Nevertheless, pulse oximetry and NIRS share the same underlying physics, which relates to the propagation of light inside the tissue as considered by the mBLL. The physics tells us that the measured optical signal (the so-called ratio-of-ratios () in pulse oximetry) depends on the extinction coefficients (’s) of oxy-hemoglobin () and deoxy-hemoglobin (Hb) (i.e., and ) and the average partial optical path length in the pulsatile arterial blood volume (). Said in another way, is the average path length that photons spend in the part of the arterial blood volume, which is expanded during the systolic phase () [in the text that follows, in Eq. (9), we will discuss an alternative interpretation based on the mBLL]. The way that these quantities are considered in relation to the measured optical signals is where pulse oximetry and NIRS differ. With pulse oximetry, a preliminary calibration is performed by fitting an equation (which may have various forms such as linear and quadratic) to the data of versus ratio-of-ratios () collected on a calibration population of healthy subjects. By carrying out this exercise, one is in fact finding coefficients that relate to , , and through empirical calibration. In Sec. 2, we derive an equation that relates to using the mBLL, similar to what was previously reported.12,13 Here, we provide a derivation of the equation and a critical description of key terms. We also point out how this equation shows the same functional dependence of on as the equations used for calibration in pulse oximetry, even though pulse oximetry equations may sometimes linearize the relationship between and . In all cases, the calibration equations used in pulse oximetry contain coefficients that depend on , , and . We hope that this derivation may be helpful to researchers in both fields of pulse oximetry and NIRS as it demonstrates the equivalence of the methods used in the two fields while also offering an opportunity to critically evaluate the meaning of the calibration coefficients used in pulse oximetry. 2.Derivation of the Relationship Between Arterial Oxygen Saturation () and Ratio-of-Ratios () Using the Modified Beer–Lambert Law (mBLL)We start our derivation by expressing the amplitude of the effective absorption coefficient change () over the entire probed volume () in terms of the absorption coefficient of arterial blood () and pulsatile arterial blood-volume change (), by using Beer’s law to relate to the oxy-hemoglobin concentration () and the deoxy-hemoglobin concentration ([Hb]) in arterial blood ( and ): Note that the subscripts represent either the red or IR wavelength, and the superscript (a) represents the arterial vascular compartment. We also observe that water absorption is neglected in Eqs. (1) and (2) for the sake of simplicity, but water absorption may be included after taking into account the water content in plasma14 and red blood cells.15 Solving the system of Eqs. (1) and (2) for and , we get Now, we write the definition of as Then, solving for in terms of , , and the ’s yields Dividing by gives: Now, we introduce the mBLL where is the change of detected intensity between two states of the medium (a baseline state and a perturbed state ), and is the change of arterial blood volume between these two states. We point out that even though the choice of baseline and perturbed states is totally arbitrary, this choice has a consequence for the interpretation of . For example, if we choose the diastolic state as baseline (i.e., ), then is the mean path length in those tissue regions where the arterioles are expanding into. On the contrary, if we choose the systolic state as baseline (i.e., ), then is the mean path length in those tissue regions where the arterioles are receding from. In both cases, we expect that the optical properties of arterial blood (related to hematocrit and ) are going to substantially affect [but may have a lesser impact on the ratio of partial path lengths of Eq. (10)]. We can connect the of Eqs. (8) and (9) to typical measurands of pulse oximetry, if we choose as the detected intensity at an intermediate state between diastolic and systolic phases [so that it becomes the direct current (DC) component in pulse oximetry] and the as the amplitude change from pulsation (i.e., is the same as the alternating current (AC) component of pulse oximetry]. Finally, the approximate equality in Eqs. (8) and (9) holds if , which is the case we consider here, even though it was noted that this approximation may not always be accurate in pulse oximetry.16,17 Continuing the derivation, next we divide the two mBLL equations [Eqs. (8) and (9)] at the two wavelengths to yieldHere, it is important to note that is exactly what is called AC/DC (i.e., pulsatile amplitude divided by average signal) in the pulse oximetry literature. Given that the ratio-of-ratios () is defined as we can rewrite Eq. (10) asFinally, we substitute Eq. (12) into Eq. (7) to get We observe that Eq. (13) is formally identical to previously reported equations for as a function of ,12,13 with the key difference that Eq. (13) contains a ratio of partial optical path lengths (’s) rather than a ratio of effective total optical path lengths through the entire tissue. This difference is crucially important because the ratio has a different value and features a different dependence on and other anatomical and biological variables than the ratio of the total optical path lengths at IR and red wavelengths, as was reported before. To arrive at a compact version of the equation, we rewrite Eq. (13) as whereEquation (14) is the key result of this derivation because it shows the relationship between and and importantly explains how the coefficients relate to ’s and ’s. Figure 1 plots versus according to Eq. (14) for various values of , considering known hemoglobin extinction coefficients (’s)19 for the optical wavelengths (’s) of 660 and 940 nm. 2.1.Assuming a Linear Relationship Between Arterial Oxygen Saturation () and Ratio-of-Ratios ()Although Eq. (14) is indeed used in pulse oximetry,16 it can also be linearized. This is common in pulse oximetry, where a linear relationship between and is assumed. To linearize Eq. (14), we use the Taylor expansion whereFor the sake of example, let , which corresponds to a range of values around 0.85. Linearizing about other values of may be done to increase the accuracy of the resulting linear equation about a certain range of ; for example, if we assume , linearizing about would result in maximum accuracy about (Fig. 1). The range of typical values of in literature depends on various factors, most notably the choice of wavelengths. However, to give an example range of values, may vary between and 1.4 when we focus on the range of 1.0 to 0.7 for using 660 and 940 nm.18 For , the Taylor expansion leads to the following linearized relationship: which we can write as whereHere, it is important to remember the definitions of , , , and [Eqs. (15)–(18)], which just depend on ’s and ’s. All ’s are known19 (given knowledge of the wavelengths), and only is unknown. For this example, we take the same ’s as in the Texas Instruments TIDA-00301 reference design18 so that and . If we assume , then and , so that Eq. (22) becomes which compares closely to the equation in the TIDA-00301 reference design,18 which isAside from the TIDA-00301 reference design, this equation is also widely used in the pulse oximetry literature20–23 and is in good agreement with another reported linearly calibrated equation from the literature that does not apply the assumption to remove the natural logarithm in Eqs. (8) and (9)17 Figure 1 visualizes the non-linearized version [Eq. (14)] for these wavelengths (i.e., 660 and 940 nm) and also plots the TIDA-00301 reference design18 equation [Eq. (26)] for comparison. This comparison shows that the TIDA-00301 reference design equation corresponds to a value of for and for . This suggests that the ratio ranges from to 0.70 when decreases from 1.0 to 0.7. This is a much smaller range of values than the one considered in Fig. 1 (0.50 to 1.00), which is the approximate range associated with the same decrease from 1.0 to 0.7 in an ideal case of a homogeneous increase in the concentration of arterial blood in the tissue. 3.DiscussionAs can be seen from Fig. 1 and the numerical example in Eq. (25), the TIDA-00301 reference design equation 18 is a linear approximation of Eq. (14) associated with a fixed value for of . When calibration is carried out for such pulse oximetry equations, all ’s and are empirically estimated. We caution thinking of a dichotomy between NIRS (i.e., mBLL-derived methods) and pulse oximetry as the underlying physics is the same. There are a few implicit assumptions in the derivation we would like to discuss. The first arises in Eq. (5), where it is assumed that all measured pulsatile changes result from blood-volume changes in the arteries alone (i.e., is the only source of intensity change ). The ubiquity and success of pulse oximetry suggest that this assumption is reasonable. Built into this assumption is that the pulsatile oscillations of and in the probed tissue volume are in phase with each other. In fact, contributions from pulsatile blood flow would result in out-of-phase oscillations of and .24 As a matter of fact, some work has shown not fully in-phase pulsatile oscillations of and ,24,25 leading us to suggest further investigation of this assumption. The second implicit assumption we would like to discuss is the introduction of in Eqs. (8) and (9). In general, optical path lengths must be introduced when using mBLL, but their meaning can be nuanced. This is one of those cases. As we isolate only pulsatile oscillations in pulse oximetry analysis, we are assuming that only comes from the vascular compartment that is associated with the pulsatile oscillation. Therefore, Eqs. (8) and (9) were introduced, in which is the average partial optical path length in the pulsatile arterial blood volume (i.e., in ). Thus, is difficult to know, and we know of no definitive way to measure it even with FD or TD NIRS. This is because knowledge of requires knowing two things: first, the spatial distribution of absolute optical properties in the tissue, and second, the spatial distribution of pulsatile arteries. This is likely different for every person who undergoes a pulse oximetry measurement; thus, we posit that the success of pulse oximetry (with calibration that needs to be constant across people) is relying on the fact that only the ratio of ’s at different wavelengths appears in Eqs. (14) and (22). With this in mind, we also suggest further investigation of this question of the variability of (or the ratio of ’s) across different subjects. This is a possible physical explanation of the racial disparities already observed in pulse oximetry methods26,27 or possible biases from blood content or finger size. Finally, we observe that is not strictly constant because it depends on , which affects the distribution of absolute optical properties in tissue. This is related to a known issue where the pulse oximetry calibration curve becomes inaccurate at low values of ,28 for which calibration adjustments as a function of have been proposed.29 The analysis presented in this work [leading to Eq. (14)] shows that the dependence of on is the key factor to consider for adjusting the calibration curve as a function of . Figure 1 suggests that the dependence of on is weaker than in an ideal case of a homogeneous increase of arterial blood volume, and more research is needed to quantify such dependence in conditions relevant to pulse oximetry. We observe that also depends on the source-detector geometry of the pulse oximeter. Source-detector geometry is different in transmittance and reflectance-based pulse oximetry. Therefore, may feature a different sensitivity to confounders in reflectance versus transmittance geometries. This suggests that investigations on how is influenced by different collection geometries may be fruitful. Finally, we would like to discuss the matter of the hemoglobin extinction coefficients (’s), especially related to the spectral features of the sources. First, let us assume that we are using laser diodes (LDs) and know the wavelength of both the red and IR sources. In this case, both Eqs. (14) and (22) have two unknowns. This requires noticing that Eq. (15) for and Eq. (17) for used in Eq. (14) only depend on ’s, which are known if source wavelengths are known19 albeit with some uncertainty.30 For this reason, we see no real advantage to Eq. (22) if we have knowledge of the source wavelengths as both Eqs. (14) and (22) have two unknowns, but Eq. (22) introduces more assumptions. Furthermore, we could go one step further to claim that Eq. (14) really only has one unknown, which is , which may simplify calibration even further. Now, we consider that most pulse oximeter devices use light-emitting diodes (LEDs), not LDs. In fact, the broad spectral features of these sources have been suggested as a reason for the observed skin-tone bias.31,32 In the case of LEDs, the discussion of calibration and above is still valid. However, ’s would be replaced with a weighted average (), weighted by the LED emission spectrum,16 which requires calibrating the LED spectrum. With all these considerations in mind, we see a great advantage in reconsidering the use of LDs in pulse oximeter devices as the advantage may outweigh the cost and safety concerns (which could be mitigated by adding a diffuse material to the LD). 4.ConclusionThe difference between conventional pulse oximetry and NIRS methods based on the mBLL lies in the consideration of , , and by the latter as opposed to empirical calibration or linearity assumptions between and , by the former. In this perspective, we have leveraged the common underlying physics of diffuse optics to link the empirical approach of pulse oximetry and the analytical description of the mBLL. The key equations are Eqs. (14) and (22), with the former coming directly from the mBLL and the latter being a linear approximation of Eq. (14). Noting that both equations have two unknowns [ and are known in Eq. (14), albeit with some uncertainty, as they are only related to extinction coefficients], we see little advantage to Eq. (22). With this in mind, we suggest that more emphasis be placed on calibrating to Eq. (14) in pulse oximetry, but we acknowledge that this would require knowledge of the source wavelengths. The value of Eq. (14) is also that it specifies the physical meaning of the four coefficients (i.e., , , , and ), thus providing specific indications on the impact of various error sources. We hope that this exercise of deriving from mBLL is informative and helpful to other researchers in both the fields of NIRS and pulse oximetry. Code and Data AvailabilityApplicable supporting code and data are available from the authors upon reasonable request. AcknowledgmentsThis work is supported by the National Institutes of Health (NIH) (Award No. R01-EB029414). G.B. would also like to acknowledge support from NIH (Award No. K12-GM133314). The content is solely the authors’ responsibility and does not necessarily represent the official views of the awarding institutions. ReferencesV. Quaresima, M. Ferrari and F. Scholkmann,
“Ninety years of pulse oximetry: history, current status, and outlook,”
J. Biomed. Opt., 29 S33307 https://doi.org/10.1117/1.JBO.29.S3.S33307 JBOPFO 1083-3668
(2024).
Google Scholar
T. Leppänen et al.,
“Pulse oximetry: the working principle, signal formation, and applications,”
Advances in the Diagnosis and Treatment of Sleep Apnea: Filling the Gap Between Physicians and Engineers, 205
–218 Springer International Publishing, Cham
(2022). Google Scholar
M. Nitzan, A. Romem and R. Koppel,
“Pulse oximetry: fundamentals and technology update,”
Med. Devices: Evidence Res., 7 231
–239 https://doi.org/10.2147/MDER.S47319
(2014).
Google Scholar
E. D. Chan, M. M. Chan, M. M. Chan,
“Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations,”
Respir. Med., 107 789
–799 https://doi.org/10.1016/j.rmed.2013.02.004 RMEDEY 0954-6111
(2013).
Google Scholar
F. Scholkmann et al.,
“A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology,”
NeuroImage, 85 6
–27 https://doi.org/10.1016/j.neuroimage.2013.05.004 NEIMEF 1053-8119
(2014).
Google Scholar
S. Fantini and A. Sassaroli,
“Frequency-domain techniques for cerebral and functional near-infrared spectroscopy,”
Front. Neurosci., 14 300 https://doi.org/10.3389/fnins.2020.00300 1662-453X
(2020).
Google Scholar
Y. Yamada, H. Suzuki and Y. Yamashita,
“Time-domain near-infrared spectroscopy and imaging: a review,”
Appl. Sci., 9 1127 https://doi.org/10.3390/app9061127
(2019).
Google Scholar
H. Ayaz et al.,
“Optical imaging and spectroscopy for the study of the human brain: status report,”
Neurophotonics, 9 S24001 https://doi.org/10.1117/1.NPh.9.S2.S24001
(2022).
Google Scholar
S. Perrey and M. Ferrari,
“Muscle oximetry in sports science: a systematic review,”
Sports Med., 48 597
–616 https://doi.org/10.1007/s40279-017-0820-1 SPMEE7 0112-1642
(2018).
Google Scholar
D. Grosenick et al.,
“Review of optical breast imaging and spectroscopy,”
J. Biomed. Opt., 21 091311 https://doi.org/10.1117/1.JBO.21.9.091311 JBOPFO 1083-3668
(2016).
Google Scholar
A. Sassaroli and S. Fantini,
“Comment on the modified Beer–Lambert law for scattering media,”
Phys. Med. Biol., 49 N255
–N257 https://doi.org/10.1088/0031-9155/49/14/N07 PHMBA7 0031-9155
(2004).
Google Scholar
A. Zourabian et al.,
“Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry,”
J. Biomed. Opt., 5 391
–405 https://doi.org/10.1117/1.1289359 JBOPFO 1083-3668
(2000).
Google Scholar
M. Nitzan et al.,
“Measurement of oxygen saturation in venous blood by dynamic near IR spectroscopy,”
J. Biomed. Opt., 5 155
–162 https://doi.org/10.1117/1.429982 JBOPFO 1083-3668
(2000).
Google Scholar
M. J. Albrink et al.,
“The displacement of serum water by the lipids of hyperlipemic serum. A new method for the rapid determination of serum water,”
J. Clin. Invest., 34
(10), 1483
–1488 https://doi.org/10.1172/JCI103199 JCINAO 0021-9738
(1955).
Google Scholar
L. J. Beilin et al.,
“The sodium, potassium, and water contents of red blood cells of healthy human adults,”
J. Clin. Invest., 45
(11), 1817
–1825 https://doi.org/10.1172/JCI105485 JCINAO 0021-9738
(1966).
Google Scholar
O. Yossef Hay et al.,
“Pulse oximetry with two infrared wavelengths without calibration in extracted arterial blood,”
Sensors, 18 3457 https://doi.org/10.3390/s18103457 SNSRES 0746-9462
(2018).
Google Scholar
N. Stubán and N. Masatsugu,
“Non-invasive calibration method for pulse oximeters,”
Period. Polytech. Electr. Eng. (Arch.), 52
(1–2), 91
–94 https://doi.org/10.3311/pp.ee.2008-1-2.11
(2008).
Google Scholar
P. Aroul,
“TIDA-00301—getting started guide—miniaturized pulse oximeter reference design,”
https://www.ti.com/lit/ug/tidu475/tidu475.pdf?ts=1725516652460
(2014).
Google Scholar
S. Prahl,
“Tabulated molar extinction coefficient for hemoglobin in water,”
https://omlc.org/spectra/hemoglobin/summary.html
(1998).
Google Scholar
P. A. Kyriacou,
“Pulse oximetry in the oesophagus,”
Physiol. Meas., 27 R1 https://doi.org/10.1088/0967-3334/27/1/R01 PMEAE3 0967-3334
(2005).
Google Scholar
Z. D. Walton et al.,
“Measuring venous oxygenation using the photoplethysmograph waveform,”
J. Clin. Monit. Comput., 24 295
–303 https://doi.org/10.1007/s10877-010-9248-y
(2010).
Google Scholar
J. P. Phillips et al.,
“Photoplethysmographic measurements from the esophagus using a new fiber-optic reflectance sensor,”
J. Biomed. Opt., 16 077005 https://doi.org/10.1117/1.3598858 JBOPFO 1083-3668
(2011).
Google Scholar
J. Koseeyaporn et al.,
“Pulse oximetry based on quadrature multiplexing of the amplitude modulated photoplethysmographic signals,”
Sensors, 23 6106 https://doi.org/10.3390/s23136106 SNSRES 0746-9462
(2023).
Google Scholar
J. M. Kainerstorfer, A. Sassaroli and S. Fantini,
“Optical oximetry of volume-oscillating vascular compartments: contributions from oscillatory blood flow,”
J. Biomed. Opt., 21 101408 https://doi.org/10.1117/1.JBO.21.10.101408 JBOPFO 1083-3668
(2016).
Google Scholar
G. Blaney et al.,
“Dual-ratio approach to pulse oximetry and the effect of skin tone,”
J. Biomed. Opt., 29 S33311 https://doi.org/10.1117/1.JBO.29.S3.S33311 JBOPFO 1083-3668
(2024).
Google Scholar
C. Shi et al.,
“The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis,”
BMC Med., 20 267 https://doi.org/10.1186/s12916-022-02452-8
(2022).
Google Scholar
W. Sjoding Michael et al.,
“Racial bias in pulse oximetry measurement,”
N. Engl. J. Med., 383 2477
–2478 https://doi.org/10.1056/NEJMc2029240
(2020).
Google Scholar
D. Wackernagel, M. Blennow and A. Hellström,
“Accuracy of pulse oximetry in preterm and term infants is insufficient to determine arterial oxygen saturation and tension,”
Acta Paediatr., 109
(11), 2251
–2257 https://doi.org/10.1111/apa.15225
(2020).
Google Scholar
J. Wu et al.,
“Self-calibrated pulse oximetry algorithm based on photon pathlength change and the application in human freedivers,”
J. Biomed. Opt., 28 115002 https://doi.org/10.1117/1.JBO.28.11.115002 JBOPFO 1083-3668
(2023).
Google Scholar
J. G. Kim and H. Liu,
“Variation of haemoglobin extinction coefficients can cause errors in the determination of haemoglobin concentration measured by near-infrared spectroscopy,”
Phys. Med. Biol., 52 6295 https://doi.org/10.1088/0031-9155/52/20/014
(2007).
Google Scholar
P. Bickler and K. K. Tremper,
“The pulse oximeter is amazing, but not perfect,”
Anesthesiology, 136 670
–671 https://doi.org/10.1097/ALN.0000000000004171 ANESAV 0003-3022
(2022).
Google Scholar
M. S. Rea and A. Bierman,
“Light source spectra are the likely cause of systematic bias in pulse oximeter readings for individuals with darker skin pigmentation,”
Br. J. Anaesth., 131 e101
–e103 https://doi.org/10.1016/j.bja.2023.04.018 BJANAD 0007-0912
(2023).
Google Scholar
BiographyGiles Blaney is a National Institutes of Health (NIH) Institutional Research and Academic Career Development Award (IRACDA) Postdoctoral Scholar in the Diffuse Optical Imaging of Tissue (DOIT) lab at Tufts University. He received his PhD from Tufts University (Medford, Massachusetts, United States) in 2022 after working in the same lab with Prof. Sergio Fantini as his advisor. Before that, he received an undergraduate degree in mechanical engineering and physics from Northeastern University (Boston, Massachusetts, United States). His current research interests include diffuse optics and its possible applications within and outside of medical imaging. Angelo Sassaroli received his PhD in physics in 2002 from the University of Electro-Communications (Tokyo, Japan). From July 2002 to August 2007, he was a research associate in the research group of Prof. Sergio Fantini at Tufts University. In September 2007, he was appointed by Tufts University as a research assistant professor. His field of research is near-infrared spectroscopy and diffuse optical tomography. Sergio Fantini is a professor of biomedical engineering and principal investigator for the DOIT Lab at Tufts University. His research activities on applying diffuse optics to biological tissues resulted in about 130 peer-reviewed scientific publications and 13 patents. He co-authored with Prof. Irving Bigio (Boston University) a textbook on “Quantitative Biomedical Optics” published by Cambridge University Press in 2016. He is a fellow of SPIE, Optica, and the American Institute for Medical and Biological Engineering (AIMBE). |