6 June 2023 Angiographic velocimetry analysis using contrast dilution gradient method with a 1000 frames per second photon-counting detector
Author Affiliations +
Abstract

Purpose

Contrast dilution gradient (CDG) analysis is a quantitative method allowing blood velocity estimation using angiographic acquisitions. Currently, CDG is restricted to peripheral vasculature due to the suboptimal temporal resolution of current imaging systems. We investigate extension of CDG methods to the flow conditions of proximal vasculature using 1000 frames per second (fps) high-speed angiographic (HSA) imaging.

Approach

We performed in-vitro HSA acquisitions using the XC-Actaeon detector and 3D-printed patient-specific phantoms. The CDG approach was used for blood velocity estimation expressed as the ratio of temporal and spatial contrast gradients. The gradients were extracted from 2D contrast intensity maps synthesized by plotting intensity profiles along the arterial centerline at each frame. In-vitro results obtained at various frame rates via temporal binning of 1000 fps data were retrospectively compared to computational fluid dynamics (CFD) velocimetry. Full-vessel velocity distributions were estimated at 1000 fps via parallel line expansion of the arterial centerline analysis.

Results

Using HSA, the CDG method displayed agreement with CFD at or above 250 fps [mean-absolute error (MAE): 2.6 ± 6.3 cm / s, p = 0.05]. Relative velocity distributions correlated well with CFD at 1000 fps with universal underapproximation due to effects of pulsatile contrast injection (MAE: 4.3 cm/s).

Conclusions

Using 1000 fps HSA, CDG-based extraction of velocities across large arteries is possible. The method is sensitive to noise; however, image processing techniques and a contrast injection, which adequately fills the vessel assist algorithm accuracy. The CDG method provides high resolution quantitative information for rapidly transient flow patterns observed in arterial circulation.

© 2023 Society of Photo-Optical Instrumentation Engineers (SPIE)
Kyle A. Williams, Allison Shields, Swetadri Vasan Setlur Nagesh, Mitchell Chudzik, Daniel R. Bednarek, Stephen Rudin, and Ciprian Ionita "Angiographic velocimetry analysis using contrast dilution gradient method with a 1000 frames per second photon-counting detector," Journal of Medical Imaging 10(3), 033502 (6 June 2023). https://doi.org/10.1117/1.JMI.10.3.033502
Received: 21 June 2022; Accepted: 19 May 2023; Published: 6 June 2023