1.IntroductionMagnetic resonance imaging (MRI) has several important advantages over computed tomography (CT) for radiation treatment planning.1 Chiefly, MRI significantly improves soft-tissue contrast over CT, which increases the accuracy and reliability of target delineation in multiple anatomic sites. A potential treatment planning process incorporating MRI as the sole imaging modality could eliminate systematic MRI-CT coregistration errors, reduce cost, minimize radiation exposure to the patient, and simplify clinical workflow.2 Despite these advantages, MRI contains neither unique nor quantitative information on electron density, which is needed for accurate dose calculations and generating reference images for patient setup. Therefore, in this proposed MRI-only-based treatment planning process, electron density information that is otherwise obtained from CT must thus be derived from the MRI by synthesizing a so-called pseudo CT (PCT).3,4 Accurate methods for estimating PCT from MRI are crucially needed. The existing estimation methods can be broadly classified into the following four categories:
Since recent dictionary-learning-based methods have not taken into consideration patient-specific features when representing an image patch, the purpose of this work is to address this issue by the following improvement.
This paper is organized as follows: we provide an overview of the proposed MRI-based PCT framework in the methods, followed by the details on anatomic signature, construction of the coupled similar patch-based dictionary via joint dictionary learning, and the prediction of target CT by sparse representation reconstruction. We compare state-of-art methods based on dictionary learning with our approach in the results and conclude that our dictionary-learning-based PCT estimation framework could be a useful tool for MRI-based radiation treatment planning for rigid structures. 2.Methods2.1.System OverviewFor a given pair of MR and CT training images, the CT image was used as the regression target of the MRI image. We assumed that the training data have been preprocessed by removing noise and uninformative regions and have been aligned and normalized using mean and standard deviation. In the training stage, for each 3-D patch of MR image, we extracted 3-D RILBP features with multiradii-level sensitivity and multiderivate image modeling including the original MR image and the derivate image filtered through standard deviation and mean filters. The mean and standard filtered images were obtained by assigning each voxel a value corresponding to the mean and standard value of the voxels in a spherical neighborhood around the voxel, respectively. The CT label of each patch’s central voxel was obtained using fuzzy C-means labeling. The most informative features corresponding to cluster the labels were identified by an LASSO operator; their statistical discriminative power was evaluated by each one’s Fisher’s score.47 For each patch of new arrival MRI, the training data were collected by searching similar neighbors within a bounded region. Then, we used a joint dictionary learning method to simultaneously train the sparse representation dictionary of both anatomic signature and CT intensities within the searched region. Last, the sparse coefficients of this patch were generated by BP46 methods under the dictionary of MRI features. Finally, we used these coefficients to reconstruct the PCT image under the well-trained dictionary of CT. A brief workflow of our method is shown in Fig. 1. 2.2.PreprocessingA rigid intrasubject registration was used to align each pair of brain CT and MR images of the same subject. All pairs were aligned onto a common space by applying a rigid-body intersubject registration. All the training pairs were further registered to a new test MR image. We performed denoising, inhomogeneity correction, and histogram matching for MR images. To reduce the computational cost of training the models, a mask was performed by removing the most useless regions, i.e., the air region outside the body contour, in all MRI images. 2.3.Feature ExtractionBased on the approach proposed by Pauly et al.,48 in which two-dimensional (2-D) LBP-like features were used for predicting organ bounding boxes in Dixon MR image sets, a 3-D rotation invariant version was implemented for this task. The 3-D version differs from 2-D version. The neighborhood voxels are not defined by a circular neighborhood but instead by a spherical one, which can be solved by the RILBP operator proposed by Fehr and Burkhardt.49 To address MRI artifacts, we used multiderivate image modeling to extract features with different image models, and also used multiple spatial scales with fixed patch size to achieve the feature’s multiradius level sensitivity as recommended previously.41 For 3-D RILBP feature extraction, we used a cube with size of and a sphere with a 3-mm radius for the first scale and then added 2 and 4 mm to those values for the second and third scales, respectively. The 3-D RILBP feature maps of multilevel and multiderivate image modeling in transverse plane are shown in Fig. 2, where each feature map’s voxel is constructed by central feature of feature vector of 3-D patch centered at that voxel’s position. 2.4.Feature SelectionWe denote as the extracted features of MR image patches centered at each position , is the set of CT voxels at each position . Recent studies50–52 have shown the potential drawback of using extremely high-dimensional features for existing learning methods. One major concern is that high-dimensional training data may contain bias and redundant components, which could over fit the learning models and in turn affect the final learning robustness. To address this challenge, we propose a feature selection method using a logistic regression function, as feature selection can be regarded as a binary regression task with respect to each dimension of the original feature.50,51 Further, the goal of feature selection is to identify and group a small subset of the most informative features. Thus, our feature selection is accomplished by implementing the sparsity constraint in the logistic regression, i.e., minimizing the following LASSO energy loss function: where denotes the original feature, is a sparse coefficient vector that denotes a binary coefficient, with nonzero denoting that the corresponding features are relevant to the anatomic classification, and zero denoting that the nonrelevant features are eliminated during the classifier learning process. denotes the intercept scalar. is an anatomical binary labeling function. is a optimization scalar and is a regularization parameter. denotes the number of voxel samples and denotes the length of vector . is computed by discriminative power, i.e., Fisher’s score47 for each feature of feature vector.The optimization tries to punish the features that have smaller discriminative power, i.e., uninformative features are eliminated from training the learning-based model. To implement this punishment, is used to be divided by . The optimal solution of intercept scalar and can be estimated by Nesterov’s53 method. Denoted as , the informative features corresponding to the nonzero entries in were selected, which have superior discriminatory power in distinguishing bone, air, and soft-tissues from each other. We defined as the MRI patch’s anatomic signature. Two-step feature selection was applied by first labeling air and nonair materials in first round and then labeling bone and soft-tissue in the second round. The air, soft-tissue, and bone labels, as shown in Fig. 3, are roughly segmented by fuzzy C-means clustering method. Structures adjacent to both bone and air are difficult to accurately disambiguate in comparison with voxels of other regions. Therefore, more samples were drawn from the bone and air regions during the feature selection process. To balance the three regional samples, ratio parameters were used to control the ratio of the number of samples drawn from air, bone, and soft-tissue regions. An example is given in Fig. 4, in which (a) and (b) show training paired MRI and CT image with two types of samples. The samples of the bone region are indicated by red asterisk whereas those selected from soft-tissue regions are indicated by green circles. Figure 4(c) shows a scattered plot of the samples corresponding to the randomly selected two features without feature selection, whereas Fig. 4(d) shows that of the samples corresponding to the two top-ranked features, which are evaluated by Fisher’s score,47 after feature selection. It can be seen with feature selection that bone regions can be distinguished from the soft tissues. 2.5.Pseudo Computed Tomography SynthesisAfter preselecting the most informative feature as an anatomic signature, we propose a joint dictionary learning framework to handle the challenge of large-scale data training. The aim was to estimate the target CT intensity from a MRI in a patch, or a group of voxels, wise local self-similarity fashion using the joint dictionary learning and sparse representation technique. If a patch (with a size of voxels) in the new arrival MR image is similar to a certain patch of training MR images, then the corresponding CT central voxel for these two MR patches were highly correlated. Each patch of the new arrival MRI image could identify several similar instances from the properly selected training MR patches via a sparse learning technique when the training data set is sufficient enough. Therefore, in this paper the correlation of MR anatomic signature and CT central voxels can be constructed by joint dictionary learning, which tracks the same sparse coefficients of both MR patches and CT intensities. To this end, the framework involves the training and predicting stages. In the training stage, the following two steps are used to efficiently learn the correlation between MR anatomic signature and CT central voxel within the selected similar templates.
The prediction stage consists of the following two steps:
2.5.1.Patch selection based on local similarityBefore constructing the coupled dictionary, patch selection is needed among all candidate MR patches to reduce the computational complexity and to enhance the prediction accuracy by excluding the irrelevant patches based on reasonable similar metrics. Specifically, the criterion for patch similarity is defined between two MRI patches and by a reflecting weight , which can be computed by calculating distance between both CT intensity and MRI anatomic signature where is the number of voxels in the target patch, is a smoothing parameter, is the standard division of all the patches and features, and denotes the balance parameter of the similarity between anatomic signature and image intensity. First, the new arrival MRI is normalized by mean and standard deviation. Then, the training set of the new MRI patch’s central position can be obtained by sorting the -nearest patches within a specific searching region where reflecting the local similarities between new MRI patch and training MRI patch . is utilized in our next step of joint dictionary learning.2.5.2.Joint dictionary learningLet denotes a database containing a set of anatomic signature, MRI patches, and corresponding CT central voxels: where denotes the anatomic signature, denotes the multidimensional space, and denotes the corresponding MRI patch and CT central voxel intensity value, respectively. To train the locally-adaptive coupled dictionary for each preselected training database, we initialized the first iteration of the dictionary with each coupled column comprising a pair of similar MRI anatomical signature and a corresponding central voxel of CT. Let denotes the set of weighted anatomic signatures of training MRI patches and denotes the set of corresponding CT intensities. The element-wise product of is obtained by Fisher’s score47 as shown in Sec. 2.4. The goal of applying omega is to further enforce the influence of the most discriminative signatures by adaptive weighting parameter. Then, the initialization of coupled dictionary is generated as . In the following iterations, supposing we have the roughly constructed dictionary of MRI signatures and of CT intensities on training data from the previous iteration, we utilize the joint dictionary learning method to simultaneously update and with the core idea that the sparse representation tracks the same sparse coefficients of both MR signatures and CT intensities. Let denotes the weights corresponding to each MRI patch computed by local self-similarity in Eq. (2), where . Then, each training MRI anatomic signature and its corresponding CT intensity can be sparsely represented by a pair of coupled and , respectively. Let denotes the set of sparse coefficients. Then the joint dictionary learning framework can be described as follows: where is the regularization parameter of column sparse term, denotes the parameter of penalization, is the joint -norm, and is ’th row of the matrix . Equation (4) can be solved by ADM method.452.5.3.PredictionFor each patch of a new arrival MRI patch, the sparse representation of a new arrival MRI patch anatomic signature under dictionary can be calculated by a BP optimization46 where denotes the anatomic signature of each new MRI patch. Finally, with the corresponding dictionary of CT , we utilized the coefficient to reconstruct the PCT intensity by a linear combination of dictionary columnsOur proposed algorithm is summarized in Algorithm 1. The default parameter setting of Algorithm 1 is shown in Table 1. Algorithm 1Pseudo CT prediction using anatomic signature and joint dictionary learning.
Table 1Default parameter setting.
3.Experiments3.1.Datasets and Quantitative MeasurementsTo test our prediction method, we applied the proposed method to 14 paired brain MR and CT images. All patients’ MRI () and CT () data were acquired using a Siemens MR and CT scanner. To quantitatively characterize the prediction accuracy, we used three widely used metrics: mean absolute error (MAE), peak signal-to-noise ratio (PSNR), normalized cross correlation (NCC), and similarity index (SI)54 for air, soft-tissue and bone regions. MAE is used to measure how close forecasts or predictions are to the actual outcomes. PSNR is an engineering term for the ratio between the maximum possible power of a signal and the power of corrupting noise that affects the fidelity of its representation. NCC is a measure of similarity between two series as a function of the displacement of one relative to the other. SI has been used to evaluate the image quality of nonrigid image registration. The three metrics are defined as where is the ground-truth CT image, is the corresponding PCT image, is the maximal intensity value of and , and is the number of voxels in the image. and are the mean of CT and PCT image, respectively. and are the standard deviation of CT and PCT image, respectively. and denote the binary mask (air, soft-tissue, and bone) of CT and PCT image, respectively. We set intensity value HU as air, HU as bone, and the rest as soft-tissue. is the number of elements in set . In general, a better prediction has lower MAE, and higher PSNR, NCC, and SI values.3.2.Parameter SettingTo compare the influence of multiple parameters within our proposed algorithm (shown in Table 1) we evaluated how variation of one parameter affects performance while fixing the other parameters to their default settings in Table 1. Our parameter testing experiment was performed by testing one patient and training the other 13 patients. This was done by comparing the three metrics in Eqs. (7)–(9). We tested the regularization parameter of Eq. (1) over the range of 0.01 to 0.1 and fixed the other parameter as default settings in Table 1. The performance was not sensitive around . The smoothing parameter and balancing parameter in Eq. (2) were empirically set to 1 and 0.5, respectively. These parameters were decided by their performance on our experiments. In Eq. (4), the parameters and were initialized by 1 and 0.5, in the following iterations, these two parameters were updated by ADM as previously recommended.45 We also fixed the patch size and tested different window sizes, and vice versa. As seen in Table 2, the best MAE and NCC were obtained with a patch size of and a window size of (Table 2). Table 2The performance with different patch size and window size.
Figures 5(a1)–5(a3) show PSNR, MAE, and NCC metrics as a function of number of similar patches, whereas the other parameters are fixed as in default setting in Table 1. From these curves, 23 similar patches are sufficient for a good prediction. However, since MAE was the highest priority metric that was optimized, the absolute change in PSNR from 23 to 25 similar patches holds less weight than MAE, which resulted in 25 similar patches being the most optimal number for prediction. The number of selected features also influences the learning performance. If most features are eliminated by feature selection, the precision of PCT prediction may be diminished due to insufficient information. Conversely, if the majority of features are selected, uninformative data may also affect the accuracy of the proposed method. To measure the performance as determined by the number of selected features, we evaluated the PSNR, NCC, and MAE metrics as shown in Figs. 5(b1)–5(b3), and fixed the other parameter as shown in Table 1. One can observe that the optimal number of selected features is 63. 3.3.Contribution of Feature Selection and Joint Dictionary LearningIn Sec. 2, we introduced two enhancements over the classic feature-based method, i.e., feature selection and joint dictionary learning. Figure 6 shows detailed visual comparisons between classic feature-based (CF), feature selection-based (FS), and feature selection with jointly dictionary learning-based (FSJDL) methods. Table 3 quantifies the above observations using 14 pairs of MR-CT leave-one-out experiments. We can visually observe the result in Fig. 6 that the FSJDL method best preserves continuity, coalition, and smoothness in the prediction results. As prediction results between methods primarily differ in regions were bone is adjacent to air, visual results within such regions are shown in Figs. 6(b2)–6(d2) and 6(b5)–6(d5). Table 4 shows the consistent improvement in FS over the CF method, and the modest improvement of FSJDL over the FS and CF methods. The prediction estimated by FS is statistically superior to the CF method for MAE, NCC, and PSNR (-test, ). Also, the prediction estimated by FSJDL is significantly superior to the FS method for MAE and NCC (). Table 3Numerical results with compared methods.
Table 4Numerical results of the proposed and state-of-the-art dictionary learning methods.
3.4.Comparison of State-of-the-Art Dictionary Learning-Based MethodsWe compared our method with two state-of-the-art dictionary-based approaches: an intensity-based method (Int)42 and a fast patch-based method (FP).43 The parameters used in these two methods were set based on their best performance. The prediction results are shown in Fig. 7. The CT images generated by our method are similar to ground truth. Our method outperforms these two methods, specifically with higher similarity in shape and lower difference values. The Int method often results in blurred predictions due to simple averaging of the similar patches. Although the FP method works better than the Int method, its prediction remains noisy. Table 4 quantifies the superiority of our method over the other two in terms of MAE, PSNR, NCC, and SI. Our method’s average MAE of 82.6 HU is significantly better than the 148.9 and 125.5 HU generated by the Int and the FP methods, respectively. Figure 8 shows the detailed prediction performance of the comparison method through histogram analyses of the entire leave-one-out template. It can be seen that for some patients, overall performance is inferior to that of others, presumably owing to lack of database diversity. 4.DiscussionIn our study, we have presented a dictionary-learning-based method of estimating a PCT image from a standard MR image. We introduced an anatomic signature based on feature selection into a dictionary-learning-based method to improve the prediction accuracy of the resultant PCT. Local self-similarity was incorporated to address the challenge of large-scale training data training. Joint dictionary learning was applied to learn the coupled dictionary of MRI anatomic signatures and CT intensities. We evaluated the clinical application of our method on MR-CT brain datasets and compared its performance with the true CT image (ground truth) and two state-of-the-art dictionary-learning-based methods. The efficacy of using 3-D RILBP was visually demonstrated in Figs. 7(d1)–7(d4) when compared with [Figs. 7(c1)–7(c4)] and numerically demonstrated by detailed metrics comparing FP and CF methods in Fig. 7. We observed that 3-D RILBP features notably improved the prediction accuracy of the PCT. Figure 2 shows that multiradii level sensitivity and multiderivate image modeling provided features containing local texture, edges and spatial information derived from a 3-D MR image patch. Thus, we concluded that 3-D RILBP features can balance image deformation invariance and sensitivity to small structural changes. The efficacy of using an anatomic signature was demonstrated via the comparison between Figs. 6(c1)–6(c6), compared with Figs. 6(b1)–6(b6), Figs. 7(e1)–7(e4) versus Figs. 7(d1)–7(d4) and Table 4. The PCT generated from FP method was superior compared with that generated with the CF method. Our results showed that the anatomic signature substantially improved the prediction accuracy, both qualitatively and quantitatively. It can be concluded that the use of anatomic signature identified by feature selection can diminish the influence of bias and redundant information contained by high-dimensional training data and preserve the informative and highly relevant features for prediction performance. Figures 6(d1)–6(d6) compared with Figs. 6(c1)–6(c6), and Figs. 7(f1)–7(f4) compared with Figs. 7(e1)–7(e4), and Fig. 8 FSJDL compared with the FS method validated the efficacy of the local self-similarity and joint dictionary learning framework. The FSJDL method was a significant improvement to FS method. Since the training was based on a small set of similar patches, a CT image could be predicted in about 2 h on 3.00 GHz Intel Xeon 8-core processor, compared to about 1 week required for competing methods in the same experimental environment. Thus, joint dictionary learning drastically reduces computation complexity. Our method was substantially better than two state-of-the-art dictionary learning methods as evaluated by the local texture, edge and spatial information anatomical signature extraction, the discriminative feature selection, and the joint dictionary learning framework. As shown in Fig. 7, our system could predict PCT images that are similar to ground-truth values and could effectively capture minute changes in electron densities. As shown in Table 4 and Fig. 8, our proposed method significantly improved the performance of PCT. The MAE between our generated PCT and ground truth CT is Hounsfield unit (HU). The previous work14 has shown that an HU error on the order of did not affect the dosimetric accuracy of intensity modulated radiation treatment planning based on PCT. This shows that PCT generated from MRI is a valid clinical application that can be applied to the current workflow of patient treatment planning. Although we have proposed an accurate method for estimating PCTs derived from brain T1-weighted MR, this work still has several limitations. First, we did not apply this method to other body sites, such as abdomen and lung. For nonrigid structures, we will have to use a deformable MRI-CT registration to deal with the intrasubject mismatch between MRI and CT. Second, a dosimetric comparison between plans generated from the simulation CT and its corresponding predicted CT has not yet been explored. Third, we did not compare our algorithm with the other state-of-the-art machine-learning-based methods. The goals here are proving the efficiency of the feature selection in manufactural feature-based dictionary learning method, and improving the recent dictionary-learning-based methods for PCT estimation. Last, we tested our algorithm on 14 patients’ brain; testing in a larger population can prove the robustness of our algorithm. 5.ConclusionHere, we proposed anatomic signature and joint dictionary learning-based methods to improve the accuracy of pseudo CT prediction. The anatomic signature was generated by identifying the most informative and discriminative features within the 3-D RFLBP features extracted from multiradii level sensitivity and multiderivate image modeling-based MR images. The feature selection was implemented by LASSO and then the Fisher’s score was adopted to evaluate the relevance of such feature anatomically and statistically. To feasibly implement the specific dictionary learning for each patch of an arriving new MRI, a bounded searching region with most similar neighbors was obtained by computing the local self-similarity. Then, a joint dictionary learning method was incorporated to simultaneously train the sparse representation coupled MRI and CT dictionaries of either MRI anatomic signature or CT intensities. Afterward, the sparse coefficients of new MRI patch signature were optimized by BP methods under the well-trained MRI dictionary. Finally, we used these coefficients to reconstruct the pseudo CT intensity under the well-trained CT dictionary. Experimental results showed that our method can accurately predict pseudo CT images in various scenarios, even in the context of large shape variation, and outperforms two state-of-the-art dictionary-learning-based methods. AcknowledgmentsThis research was supported in part by the National Cancer Institute of the National Institutes of Health under Award No. R01CA215718, the Department of Defense (DoD) Prostate Cancer Research Program (PCRP) Award No. W81XWH-13-1-0269 and Dunwoody Golf Club Prostate Cancer Research Award, a philanthropic award provided by the Winship Cancer Institute of Emory University. ReferencesJ. M. Edmund and T. Nyholm,
“A review of substitute CT generation for MRI-only radiation therapy,”
Radiat. Oncol., 12 28
(2017). https://doi.org/10.1186/s13014-016-0747-y Google Scholar
S. Devic,
“MRI simulation for radiotherapy treatment planning,”
Med. Phys., 39
(11), 6701
–6711
(2012). https://doi.org/10.1118/1.4758068 MPHYA6 0094-2405 Google Scholar
D. Andreasen et al.,
“Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain,”
Med. Phys., 42
(4), 1596
–1605
(2015). https://doi.org/10.1118/1.4914158 MPHYA6 0094-2405 Google Scholar
M. Kapanen and M. Tenhunen,
“T1/T2*-weighted MRI provides clinically relevant pseudo-CT density data for the pelvic bones in MRI-only based radiotherapy treatment planning,”
Acta Oncol., 52
(3), 612
–618
(2013). https://doi.org/10.3109/0284186X.2012.692883 Google Scholar
J. Uha et al.,
“MRI-based treatment planning with pseudo CT generated through atlas registration,”
Med. Phys., 41
(5), 051711
(2014). https://doi.org/10.1118/1.4873315 MPHYA6 0094-2405 Google Scholar
J. Sjolund et al.,
“Generating patient specific pseudo-CT of the head from MR using atlas-based regression,”
Phys. Med. Biol., 60
(2), 825
–839
(2015). https://doi.org/10.1088/0031-9155/60/2/825 PHMBA7 0031-9155 Google Scholar
E. Schreibmann et al.,
“MR-based attenuation correction for hybrid PET-MR brain imaging systems using deformable image registration,”
Med. Phys., 37
(5), 2101
–2109
(2010). https://doi.org/10.1118/1.3377774 MPHYA6 0094-2405 Google Scholar
J. A. Dowling et al.,
“An atlas-based electron density mapping method for magnetic resonance imaging (MRI)-alone treatment planning and adaptive MRI-based prostate radiation therapy,”
Int. J. Radiat. Oncol., 83
(1), E5
–E11
(2012). https://doi.org/10.1016/j.ijrobp.2011.11.056 Google Scholar
J. Sjolund et al.,
“Skull segmentation in MRI by a support vector machine combining local and global features,”
in 22nd Int. Conf. on Pattern Recognition,
3274
–3279
(2014). https://doi.org/10.1109/ICPR.2014.564 Google Scholar
M. Hofmann et al.,
“MRI-based attenuation correction for PET/MRI: a novel approach combining pattern recognition and atlas registration,”
J. Nucl. Med., 49
(11), 1875
–1883
(2008). https://doi.org/10.2967/jnumed.107.049353 JNMEAQ 0161-5505 Google Scholar
P. Khateri et al.,
“A novel segmentation approach for implementation of MRAC in head PET/MRI employing short-TE MRI and 2-point Dixon method in a fuzzy C-means framework,”
Nucl. Instrum. Methods Phys. Res. Sect. A, 734 171
–174
(2014). https://doi.org/10.1016/j.nima.2013.09.006 Google Scholar
P. Khateri et al.,
“Generation of a four-class attenuation map for MRI-based attenuation correction of PET Data in the head area using a novel combination of STE/Dixon-MRI and FCM clustering,”
Mol. Imaging Biol., 17
(6), 884
–892
(2015). https://doi.org/10.1007/s11307-015-0849-1 Google Scholar
H. Yu et al.,
“Toward magnetic resonance-only simulation: segmentation of bone in MR for radiation therapy verification of the head,”
Int. J. Radiat. Oncol., 89
(3), 649
–657
(2014). https://doi.org/10.1016/j.ijrobp.2014.03.028 Google Scholar
M. S. R. Gudur et al.,
“A unifying probabilistic Bayesian approach to derive electron density from MRI for radiation therapy treatment planning,”
Phys. Med. Biol., 59
(21), 6595
–6606
(2014). https://doi.org/10.1088/0031-9155/59/21/6595 PHMBA7 0031-9155 Google Scholar
C. Siversson et al.,
“Technical note: MRI only prostate radiotherapy planning using the statistical decomposition algorithm,”
Med. Phys., 42
(10), 6090
–6097
(2015). https://doi.org/10.1118/1.4931417 MPHYA6 0094-2405 Google Scholar
X. F. Yang and B. W. Fei,
“Multiscale segmentation of the skull in MR images for MRI-based attenuation correction of combined MR/PET,”
J. Am. Med. Inf. Assoc., 20
(6), 1037
–1045
(2013). https://doi.org/10.1136/amiajnl-2012-001544 Google Scholar
A. P. Aitken et al.,
“Improved UTE-based attenuation correction for cranial PET-MR using dynamic magnetic field monitoring,”
Med. Phys., 41
(1), 012302
(2014). https://doi.org/10.1118/1.4837315 MPHYA6 0094-2405 Google Scholar
C. Buerger et al.,
“Investigation of MR-based attenuation correction and motion compensation for hybrid PET/MR,”
IEEE Trans. Nucl. Sci., 59
(5), 1967
–1976
(2012). https://doi.org/10.1109/TNS.2012.2209127 IETNAE 0018-9499 Google Scholar
J. M. Edmund et al.,
“A voxel-based investigation for MRI-only radiotherapy of the brain using ultra short echo times,”
Phys. Med. Biol., 59
(23), 7501
–7519
(2014). https://doi.org/10.1088/0031-9155/59/23/7501 PHMBA7 0031-9155 Google Scholar
V. Keereman et al.,
“MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences,”
J. Nucl. Med., 51
(5), 812
–818
(2010). https://doi.org/10.2967/jnumed.109.065425 JNMEAQ 0161-5505 Google Scholar
L. B. Aasheim et al.,
“PET/MR brain imaging: evaluation of clinical UTE-based attenuation correction,”
Eur. J. Nucl. Med. Mol. Imaging, 42
(9), 1439
–1446
(2015). https://doi.org/10.1007/s00259-015-3060-3 Google Scholar
J. Cabello et al.,
“MR-based attenuation correction using ultrashort-echo-time pulse sequences in dementia patients,”
J. Nucl. Med., 56
(3), 423
–429
(2015). https://doi.org/10.2967/jnumed.114.146308 JNMEAQ 0161-5505 Google Scholar
Y. Berker et al.,
“MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/dixon MRI sequence,”
J. Nucl. Med., 53
(9), 796
–804
(2012). https://doi.org/10.2967/jnumed.111.092577 JNMEAQ 0161-5505 Google Scholar
C. N. Ladefoged et al.,
“Region specific optimization of continuous linear attenuation coefficients based on UTE (RESOLUTE): application to PET/MR brain imaging,”
Phys. Med. Biol., 60
(20), 8047
–8065
(2015). https://doi.org/10.1088/0031-9155/60/20/8047 PHMBA7 0031-9155 Google Scholar
S. H. Hsu et al.,
“Investigation of a method for generating synthetic CT models from MRI scans of the head and neck for radiation therapy,”
Phys. Med. Biol., 58
(23), 8419
–8435
(2013). https://doi.org/10.1088/0031-9155/58/23/8419 PHMBA7 0031-9155 Google Scholar
C. M. Rank et al.,
“MRI-based treatment plan simulation and adaptation for ion radiotherapy using a classification-based approach,”
Radiat. Oncol., 8 51
(2013). https://doi.org/10.1186/1748-717X-8-51 Google Scholar
C. M. Rank et al.,
“MRI-based simulation of treatment plans for ion radiotherapy in the brain region,”
Radiother. Oncol., 109
(3), 414
–418
(2013). https://doi.org/10.1016/j.radonc.2013.10.034 RAONDT 0167-8140 Google Scholar
J. Kim et al.,
“Implementation of a novel algorithm for generating synthetic CT images from magnetic resonance imaging data sets for prostate cancer radiation therapy,”
Int. J. Radiat. Oncol., 91
(1), 39
–47
(2015). https://doi.org/10.1016/j.ijrobp.2014.09.015 Google Scholar
J. Korhonen et al.,
“Influence of MRI-based bone outline definition errors on external radiotherapy dose calculation accuracy in heterogeneous pseudo-CT images of prostate cancer patients,”
Acta Oncol., 53
(8), 1100
–1106
(2014). https://doi.org/10.3109/0284186X.2014.929737 Google Scholar
J. Korhonen et al.,
“A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer,”
Med. Phys., 41
(1), 011704
(2014). https://doi.org/10.1118/1.4842575 MPHYA6 0094-2405 Google Scholar
X. F. Yang and B. W. Fei,
“A skull segmentation method for brain MR images based on multiscale bilateral filtering scheme,”
Proc. SPIE, 7623 76233K
(2010). https://doi.org/10.1117/12.844677 PSISDG 0277-786X Google Scholar
B. K. Navalpakkam et al.,
“Magnetic resonance-based attenuation correction for PET/MR hybrid imaging using continuous valued attenuation maps,”
Invest. Radiol., 48
(5), 323
–332
(2013). https://doi.org/10.1097/RLI.0b013e318283292f INVRAV 0020-9996 Google Scholar
A. Johansson, M. Karlsson and T. Nyholm,
“CT substitute derived from MRI sequences with ultrashort echo time,”
Med. Phys., 38
(5), 2708
–2714
(2011). https://doi.org/10.1118/1.3578928 MPHYA6 0094-2405 Google Scholar
A. Johanson et al.,
“Improved quality of computed tomography substitute derived from magnetic resonance (MR) data by incorporation of spatial information—potential application for MR-only radiotherapy and attenuation correction in positron emission tomography,”
Acta Oncol., 52
(7), 1369
–1373
(2013). https://doi.org/10.3109/0284186X.2013.819119 Google Scholar
J. H. Jonsson et al.,
“Treatment planning of intracranial targets on MRI derived substitute CT data,”
Radiother. Oncol., 108
(1), 118
–122
(2013). https://doi.org/10.1016/j.radonc.2013.04.028 RAONDT 0167-8140 Google Scholar
J. H. Jonsson et al.,
“Accuracy of inverse treatment planning on substitute CT images derived from MR data for brain lesions,”
Radiat. Oncol., 10 13
(2015). https://doi.org/10.1186/s13014-014-0308-1 Google Scholar
R. J. Li et al.,
“Deep learning based imaging data completion for improved brain disease diagnosis,”
Lect. Notes Comput. Sci., 8675 305
–312
(2014). https://doi.org/10.1007/978-3-319-10443-0 LNCSD9 0302-9743 Google Scholar
A. Jog, A. Carass and J. L. Prince,
“Improving magnetic resonance resolution with supervised learning,”
in IEEE 11th Int. Symp. on Biomedical Imaging (ISBI),
987
–990
(2014). https://doi.org/10.1109/ISBI.2014.6868038 Google Scholar
D. Andreasen et al.,
“Computed tomography synthesis from magnetic resonance images in the pelvis using multiple random forests and auto-context features,”
Proc. SPIE, 9784 978417
(2016). https://doi.org/10.1117/12.2216924 PSISDG 0277-786X Google Scholar
S. L. S. Chan et al.,
“Automated classification of bone and air volumes for hybrid PET-MRI brain imaging,”
in Int. Conf. on Digital Image Computing: Techniques and Applications (Dicta),
1
–8
(2013). https://doi.org/10.1109/DICTA.2013.6691483 Google Scholar
T. Huynh et al.,
“Estimating CT image from MRI data using structured random forest and auto-context model,”
IEEE Trans. Med. Imaging, 35
(1), 174
–183
(2016). https://doi.org/10.1109/TMI.2015.2461533 ITMID4 0278-0062 Google Scholar
D. Andreasen, K. Van Leemput and J. M. Edmund,
“A patch-based pseudo-CT approach for MRI-only radiotherapy in the pelvis,”
Med. Phys., 43
(8), 4742
–4752
(2016). https://doi.org/10.1118/1.4958676 MPHYA6 0094-2405 Google Scholar
A. Torrado-Carvajal et al.,
“Fast patch-based pseudo-CT synthesis from T1-weighted MR images for PET/MR attenuation correction in brain studies,”
J. Nucl. Med., 57
(1), 136
–143
(2016). https://doi.org/10.2967/jnumed.115.156299 JNMEAQ 0161-5505 Google Scholar
S. Aouadi et al.,
“Sparse patch-based method applied to MRI-only radiotherapy planning,”
Phys. Med., 32 309
(2016). https://doi.org/10.1016/j.ejmp.2016.07.173 Google Scholar
S. Boyd et al.,
“Distributed optimization and statistical learning via the alternating direction method of multipliers,”
Found. Trends Mach. Learn., 3
(1), 1
–122
(2010). https://doi.org/10.1561/2200000016 Google Scholar
W. Deng, W. Yin and Y. Zhang,
“Group sparse optimization by alternating direction method,”
Proc. SPIE, 8858 88580R
(2013). https://doi.org/10.1117/12.2024410 PSISDG 0277-786X Google Scholar
M. W. Ayech and D. Ziou,
“Automated feature weighting and random pixel sampling in k-means clustering for Terahertz image segmentation,”
in IEEE Conf. on Computer Vision and Pattern Recognition Workshops (CVPRW),
(2015). https://doi.org/10.1109/CVPRW.2015.7301294 Google Scholar
O. Pauly et al.,
“Fast multiple organ detection and localization in whole-body MR dixon sequences,”
Lect. Notes Comput. Sci., 6893 239
–247
(2011). https://doi.org/10.1007/978-3-642-23626-6 LNCSD9 0302-9743 Google Scholar
J. Fehr and H. Burkhardt,
“3D rotation invariant local binary patterns,”
in 19th Int. Conf. on Pattern Recognition,
616
–619
(2008). https://doi.org/10.1109/ICPR.2008.4761098 Google Scholar
M. Avalos et al.,
“Sparse conditional logistic regression for analyzing large-scale matched data from epidemiological studies: a simple algorithm,”
BMC Bioinf., 16 S1
(2015). https://doi.org/10.1186/1471-2105-16-S6-S1 BBMIC4 1471-2105 Google Scholar
S. Aseervatham et al.,
“A sparse version of the ridge logistic regression for large-scale text categorization,”
Pattern Recognit. Lett., 32
(2), 101
–106
(2011). https://doi.org/10.1016/j.patrec.2010.09.023 PRLEDG 0167-8655 Google Scholar
M. Yaqub et al.,
“Investigation of the role of feature selection and weighted voting in random forests for 3-D volumetric segmentation,”
IEEE Trans. Med. Imaging, 33
(2), 258
–271
(2014). https://doi.org/10.1109/TMI.2013.2284025 ITMID4 0278-0062 Google Scholar
Y. Nesterov, Introductory Lectures on Convex Optimization: A Basic Course, Springer US, New York
(2003). Google Scholar
F. H. Tang and H. S. H. Ip,
“Image fusion enhancement of deformable human structures using a two-stage warping-deformable strategy: a content-based image retrieval consideration,”
Inf. Syst. Front., 11
(4), 381
–389
(2009). https://doi.org/10.1007/s10796-009-9151-6 Google Scholar
|