26 October 2022 Pre-deployment assessment of an AI model to assist radiologists in chest X-ray detection and identification of lead-less implanted electronic devices for pre-MRI safety screening: realized implementation needs and proposed operational solutions
Richard D. White, Mutlu Demirer, Vikash Gupta, Ronnie A. Sebro, Frederick M. Kusumoto, Barbaros S. Erdal
Author Affiliations +
Abstract

Purpose

Chest X-ray (CXR) use in pre-MRI safety screening, such as for lead-less implanted electronic device (LLIED) recognition, is common. To assist CXR interpretation, we “pre-deployed” an artificial intelligence (AI) model to assess (1) accuracies in LLIED-type (and consequently safety-level) identification, (2) safety implications of LLIED nondetections or misidentifications, (3) infrastructural or workflow requirements, and (4) demands related to model adaptation to real-world conditions.

Approach

A two-tier cascading methodology for LLIED detection/localization and identification on a frontal CXR was applied to evaluate the performance of the original nine-class AI model. With the unexpected early appearance of LLIED types during simulated real-world trialing, retraining of a newer 12-class version preceded retrialing. A zero footprint (ZF) graphical user interface (GUI)/viewer with DICOM-based output was developed for inference-result display and adjudication, supporting end-user engagement and model continuous learning and/or modernization.

Results

During model testing or trialing using both the nine-class and 12-class models, robust detection/localization was consistently 100%, with mAP 0.99 from fivefold cross-validation. Safety-level categorization was high during both testing (AUC ≥ 0.98 and ≥0.99, respectively) and trialing (accuracy 98% and 97%, respectively). LLIED-type identifications by the two models during testing (1) were 98.9% and 99.5% overall correct and (2) consistently showed AUC ≥ 0.92 (1.00 for 8/9 and 9/12 LLIED-types, respectively). Pre-deployment trialing of both models demonstrated overall type-identification accuracies of 94.5% and 95%, respectively. Of the small number of misidentifications, none involved MRI-stringently conditional or MRI-unsafe types of LLIEDs. Optimized ZF GUI/viewer operations led to greater user-friendliness for radiologist engagement.

Conclusions

Our LLIED-related AI methodology supports (1) 100% detection sensitivity, (2) high identification (including MRI-safety) accuracy, and (3) future model deployment with facilitated inference-result display and adjudication for ongoing model adaptation to future real-world experiences.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Richard D. White, Mutlu Demirer, Vikash Gupta, Ronnie A. Sebro, Frederick M. Kusumoto, and Barbaros S. Erdal "Pre-deployment assessment of an AI model to assist radiologists in chest X-ray detection and identification of lead-less implanted electronic devices for pre-MRI safety screening: realized implementation needs and proposed operational solutions," Journal of Medical Imaging 9(5), 054504 (26 October 2022). https://doi.org/10.1117/1.JMI.9.5.054504
Received: 10 December 2021; Accepted: 23 September 2022; Published: 26 October 2022