1 May 2011 Vacuum wavelength calibration of frequency-stabilized He-Ne lasers used in commercial laser
Won-Kyu Lee, Ho Suhng Suh, Chu Shik Kang
Author Affiliations +
Abstract
We report on the calibration results of the wavelength of the lasers used in commercial laser interferometers, including the vacuum wavelength deviation from its nominal value, the vacuum wavelength stability, the repeatability of stabilized wavelength, and the secular change of these properties, by analyzing the calibration data accumulated for more than two decades. We present an experimental method of calibrating the frequency of laser interferometers. We also propose a proper time interval and an appropriate uncertainty expression for the vacuum wavelength calibration of the laser interferometers.
©(2011) Society of Photo-Optical Instrumentation Engineers (SPIE)
Won-Kyu Lee, Ho Suhng Suh, and Chu Shik Kang "Vacuum wavelength calibration of frequency-stabilized He-Ne lasers used in commercial laser," Optical Engineering 50(5), 054301 (1 May 2011). https://doi.org/10.1117/1.3570680
Published: 1 May 2011
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.