Recently, digital mammography with a photon counting silicon detector has been developed. With the aim of reducing the exposure dose, we have proposed a new mammography system that uses a cadmium telluride series photon counting detector. In addition, we also propose to use a high energy X-ray spectrum with a tungsten anode. The purpose of this study was assessed that the effectiveness of the high X-ray energy spectrum in terms of image quality using a Monte Carlo simulation. The proposed photon counting system with the high energy X-ray is compared to a conventional flat panel detector system with a Mo/Rh spectrum. The contrast-to-noise ratio (CNR) is calculated from simulation images with the use of breast phantoms. The breast model phantoms differed by glandularity and thickness, which were determined from Japanese clinical mammograms. We found that the CNR values were higher in the proposed system than in the conventional system. The number of photons incident on the detector was larger in the proposed system, so that the noise values was lower in comparison with the conventional system. Therefore, the high energy spectrum yielded the same CNR as using the conventional spectrum while allowing a considerable dose reduction to the breast.
KEYWORDS: Sensors, X-rays, Mammography, Polymethylmethacrylate, X-ray imaging, Monte Carlo methods, Modulation transfer functions, Prototyping, Image quality, Imaging systems
We propose a new mammography system using a cadmium telluride (CdTe) photon-counting detector for exposure dose
reduction. In contrast to conventional mammography, this system uses high-energy X-rays. This study evaluates the
usefulness of this system in terms of the absorbed dose distribution and contrast-to-noise ratio (CNR) at acrylic step using
a Monte Carlo simulation. In addition, we created a prototype system that uses a CdTe detector and automatic movement
stage. For various conditions, we measured the properties and evaluated the quality of images produced by the system. The
simulation result for a tube voltage of 40 kV and tungsten/barium (W/Ba) as a target/filter shows that the surface dose was
reduced more than 60% compared to that under conventional conditions. The CNR of our proposal system also became
higher than that under conventional conditions. The point at which the CNRs coincide for 4 cm polymethyl methacrylate
(PMMA) at the 2-mm-thick step corresponds to a dose reduction of 30%, and these differences increased with increasing
phantom thickness. To improve the image quality, we determined the problematic aspects of the scanning system. The
results of this study indicate that, by using a higher X-ray energy than in conventional mammography, it is possible to
obtain a significant exposure dose reduction without loss of image quality. Further, the image quality of the prototype
system can be improved by optimizing the balance between the shift-and-add operation and the output of the X-ray tube. In
future work, we will further examine these improvement points.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.