10 April 2023 Correlation-based ultrasound imaging using experimental baseline
Author Affiliations +
Abstract
This paper describes a novel way to exploit the correlation-based (CB) imaging algorithm Excitelet, based on correlation of measured signals with reference signals. In this study, reference signals are approximated using an experimental baseline instead of computing it. The reference signals are obtained by experimentally measuring the Full-Matrix Capture (FMC) with individual reflectors located at one point at a time of a determined imaging grid. This experimental CB (Exp CB) approach makes fewer model assumptions than the numerical CB (Num CB) method, meaning that the baseline is closer to reality, resulting in better spatial resolution and contrast of the imaging.
Conference Presentation
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tamara Krpic, Maxime Bilodeau, Nicolas Quaegebeur, and Patrice Masson "Correlation-based ultrasound imaging using experimental baseline", Proc. SPIE 12470, Medical Imaging 2023: Ultrasonic Imaging and Tomography, 1247008 (10 April 2023); https://doi.org/10.1117/12.2654218
Advertisement
Advertisement
KEYWORDS
Ultrasonography

Nondestructive evaluation

Medical imaging applications

Reflectors

Spatial resolution

Back to Top