1 August 1991 Enhancements to the radiometric calibration facility for the Clouds and the Earth's Radiant Energy System instruments
Author Affiliations +
Abstract
Enhancements which have been made in the Radiometric Calibration Facility (RCF) of the CERES payload are described. These include narrow field blackbody and wide field of view blackbody sources, an active cavity radiometer, and a shortwave reference source. They permit the RCF to calibrate the CERES instruments to better than +/- 0.5 percent absolute radiometric accuracy in the 3.5 to above 50 micron wavelength region and to +/- 1.0 percent in the 0.3-3.5 micron region.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mark A. Folkman, Peter J. Jarecke, and Lane A. Darnton "Enhancements to the radiometric calibration facility for the Clouds and the Earth's Radiant Energy System instruments", Proc. SPIE 1493, Calibration of Passive Remote Observing Optical and Microwave Instrumentation, (1 August 1991); https://doi.org/10.1117/12.46703