21 November 2000 Characterization of a large-volume multi-element CdZnTe detector
Thomas H. Prettyman, Michael C. Browne, Kiril D. Ianakiev, Calvin E. Moss, Stephen E. Soldner
Author Affiliations +
Abstract
In this paper, we present results of experiments to characterize a large-volume (4 cm3), multi-element CdZnTe detector for high-efficiency, gamma-ray spectroscopy. The module includes an array of eight 0.5 cm3 coplanar grid detectors manufactured by eV Products. An eight-channel data acquisition system with list mode output is used to record gamma-ray events for each detector in the array. The list mode data are analyzed to determine the efficiency for coincidence events and to demonstrate different modes of operation (e.g., Compton suppression). The total efficiency of the array is found to match Monte Carlo calculations to within a few percent; however, the full-energy (photopeak) efficiency is significantly lower than predicted by Monte Carlo. The observed difference is probably caused by a combination of electrode design and charge transport properties. Approaches to improve full-energy efficiency are proposed.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Thomas H. Prettyman, Michael C. Browne, Kiril D. Ianakiev, Calvin E. Moss, and Stephen E. Soldner "Characterization of a large-volume multi-element CdZnTe detector", Proc. SPIE 4141, Hard X-Ray, Gamma-Ray, and Neutron Detector Physics II, (21 November 2000); https://doi.org/10.1117/12.407567
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Gamma radiation

Monte Carlo methods

Electrodes

Multi-element lenses

Data acquisition

Crystals

RELATED CONTENT

Strontium iodide radiation instrumentation II (SIRI-2)
Proceedings of SPIE (September 09 2019)
Multi-element neutron energy spectrometer
Proceedings of SPIE (September 11 2009)
Field-deployable gamma-radiation detectors for DHS use
Proceedings of SPIE (September 24 2007)
Photonic Data Recorder Development Update
Proceedings of SPIE (March 11 1987)

Back to Top