Paper
30 January 2012 Automated detection of chorio-scleral interface using polarization-sensitive optical coherence tomography
Author Affiliations +
Abstract
A polarization sensitive optical coherence tomography based automated algorithm for segmentation of the chorio-scleral interface is presented. The algorithm employs a two-step segmentation approach. At first, local birefringence based segmentation with low precision is performed to roughly distinguish the choroid and sclera. Successively, a depth oriented slope fitting to phase retardation is applied in both the choroid and sclera. The interface is determined as the cross-point of the two phase retardation slope lines. The algorithm shows potential for functional, objective, and volumetric choroid thickness measurement.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lian Duan, Masahiro Yamanari, and Yoshiaki Yasuno "Automated detection of chorio-scleral interface using polarization-sensitive optical coherence tomography", Proc. SPIE 8213, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XVI, 82131W (30 January 2012); https://doi.org/10.1117/12.907662