1 November 2011 Development and performance of a filter radiometer monitor system for integrating sphere sources
Author Affiliations +
Abstract
The NASA Goddard Space Flight Center (GSFC) Radiometric Calibration Laboratory (RCL) maintains several large integrating sphere sources covering the visible to the shortwave infrared wavelength range. Two critical, functional requirements of an integrating sphere source are short- and long-term operational stability and repeatability. Monitoring the source is essential in determining the origin of systemic errors, thus increasing confidence in source performance and quantifying repeatability. If monitor data falls outside the established parameters, this could be an indication that the source requires maintenance or recalibration against the National Institute of Science and Technology irradiance standard. The GSFC RCL has developed a Filter Radiometer Monitoring System (FRMS) to continuously monitor the performance of its integrating sphere calibration sources in the 400 to 2400 nm region. Sphere output change mechanisms include lamp aging, coating (e.g., BaSO4) deterioration, and ambient water vapor level. The FRMS wavelength bands are selected to quantify changes caused by these mechanisms. The FRMS design and operation are presented, as well as data from monitoring four of the RCL's integrating sphere sources.
©(2011) Society of Photo-Optical Instrumentation Engineers (SPIE)
Leibo Ding, John W. Cooper, Gilbert R. Smith, Matthew G. Kowalewski, Robert A. Barnes, Eugene Waluschka, and James J. Butler "Development and performance of a filter radiometer monitor system for integrating sphere sources," Optical Engineering 50(11), 113603 (1 November 2011). https://doi.org/10.1117/1.3646532
Published: 1 November 2011
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Optical spheres

Phase modulation

Integrating spheres

Radiometry

Lamps

Calibration

Back to Top