This work presents the prototype design and the status of the project.
Metis features two channels to image the solar corona in two different spectral bands: in the HI Lyman ∝ at 121.6 nm, and in the polarized visible light band (580 – 640 nm). Metis is a solar coronagraph adopting an “inverted occulted” configuration. The inverted external occulter (IEO) is a circular aperture followed by a spherical mirror which back rejects the disk light. The reflected disk light exits the instrument through the IEO aperture itself, while the passing coronal light is collected by the Metis telescope. Common to both channels, the Gregorian on-axis telescope is centrally occulted and both the primary and the secondary mirror have annular shape.
Classic alignment methods adopted for on-axis telescope cannot be used, since the on-axis field is not available. A novel and ad hoc alignment set-up has been developed for the telescope alignment.
An auxiliary visible optical ground support equipment source has been conceived for the telescope alignment. It is made up by four collimated beams inclined and dimensioned to illuminate different sections of the annular primary mirror without being vignetted by other optical or mechanical elements of the instrument.
The entire alignment and verification phase has been performed by the Metis team in collaboration with Thales Alenia Space Torino and took place in ALTEC (Turin) at the Optical Payload System Facility using the Space Optics Calibration Chamber infrastructure, a vacuum chamber especially built and tested for the alignment and calibration of the Metis coronagraph, and suitable for tests of future payloads.
The goal of the alignment, integration, verification and calibration processes is to measure the parameters of the telescope, and the characteristics of the two Metis channels: visible and ultraviolet. They work in parallel thanks to the peculiar optical layout. The focusing and alignment performance of the two channels must be well understood, and the results need to be easily compared to the requirements. For this, a dedicated illumination method, with both channels fed by the same source, has been developed; and a procedure to perform a simultaneous through focus analysis has been adopted.
In this paper the final optical performance achieved by Metis is reported and commented.
The stray light calibration was performed in a clean environment in front of the OPSys solar disk divergence simulator (at ALTEC, in Torino, Italy), which is able to emulate different heliocentric distances. Ground calibrations were a unique opportunity to map the Metis stray light level thanks to a pure solar disk simulator without the solar corona. The stray light calibration was limited to the visible light case, being the most stringent. This work is focused on the description of the laboratory facility that was used to perform the stray light calibration and on the calibration results.